首页> 外文期刊>Ultrasonics, Ferroelectrics and Frequency Control, IEEE Transactions on >2-D high-frame-rate dynamic elastography using delay compensated and angularly compounded motion vectors: Preliminary results
【24h】

2-D high-frame-rate dynamic elastography using delay compensated and angularly compounded motion vectors: Preliminary results

机译:使用延迟补偿和角度复合运动矢量的二维高帧率动态弹性成像:初步结果

获取原文
获取原文并翻译 | 示例

摘要

This paper describes a new ultrasound-based system for high-frame-rate measurement of periodic motion in 2-D for tissue elasticity imaging. Similarly to conventional 2-D flow vector imaging, the system acquires the RF signals from the region of interest at multiple steering angles. A custom sector subdivision technique is used to increase the temporal resolution while keeping the total acquisition time within the range suitable for real-time applications. Within each sector, 1-D motion is estimated along the beam direction. The intraand inter-sector delays are compensated using our recently introduced delay compensation algorithm. In-plane 2-D motion vectors are then reconstructed from these delay-compensated 1-D motions. We show that Young's modulus images can be reconstructed from these 2-D motion vectors using local inversion algorithms. The performance of the system is validated quantitatively using a commercial flow phantom and a commercial elasticity phantom. At the frame rate of 1667 Hz, the estimated flow velocities with the system are in agreement with the velocity measured with a pulsed-wave Doppler imaging mode of a commercial ultrasound machine with manual angle correction. At the frame rate of 1250 Hz, phantom Young¿s moduli of 29, 6, and 54 kPa for the background, the soft inclusion, and the hard inclusion, are estimated to be 30, 11, and 53 kPa, respectively.
机译:本文介绍了一种新的基于超声的系统,用于以高帧速测量二维运动的组织弹性成像。与常规2D流矢量成像类似,系统以多个转向角从感兴趣区域获取RF信号。使用自定义扇区细分技术可提高时间分辨率,同时将总采集时间保持在适合实时应用的范围内。在每个扇区内,沿波束方向估计一维运动。使用我们最近推出的延迟补偿算法可以补偿扇区内和扇区间的延迟。然后从这些延迟补偿的一维运动中重建平面内二维运动矢量。我们表明,可以使用局部反演算法从这些二维运动向量中重构杨氏模量图像。使用商业流动模型和商业弹性模型对系统的性能进行定量验证。在1667 Hz的帧频下,系统的估计流速与使用手动角度校正的商用超声仪的脉冲多普勒成像模式测得的速度一致。在1250 Hz的帧频下,背景,软包含和硬包含的幻影杨氏模量分别为29、6和54 kPa,估计分别为30、11和53 kPa。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号