首页> 外文期刊>IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control >Accurate Angle Estimator for High-Frame-Rate 2-D Vector Flow Imaging
【24h】

Accurate Angle Estimator for High-Frame-Rate 2-D Vector Flow Imaging

机译:高帧率二维矢量流成像的精确角度估计器

获取原文
获取原文并翻译 | 示例

摘要

This paper presents a novel approach for estimating 2-D flow angles using a high-frame-rate ultrasound method. The angle estimator features high accuracy and low standard deviation (SD) over the full 360° range. The method is validated on Field II simulations and phantom measurements using the experimental ultrasound scanner SARUS and a flow rig before being tested in vivo. An 8-MHz linear array transducer is used with defocused beam emissions. In the simulations of a spinning disk phantom, a 360° uniform behavior on the angle estimation is observed with a median angle bias of 1.01° and a median angle SD of 1.8°. Similar results are obtained on a straight vessel for both simulations and measurements, where the obtained angle biases are below 1.5° with SDs around 1°. Estimated velocity magnitudes are also kept under 10% bias and 5% relative SD in both simulations and measurements. An in vivo measurement is performed on a carotid bifurcation of a healthy individual. A 3-s acquisition during three heart cycles is captured. A consistent and repetitive vortex is observed in the carotid bulb during systoles.
机译:本文提出了一种使用高帧率超声方法估计二维流角的新颖方法。角度估算器在整个360°范围内具有高精度和低标准偏差(SD)。在进行体内测试之前,使用实验超声扫描仪SARUS和流动装置对Field II模拟和体模测量进行了验证。使用8 MHz线性阵列换能器以散焦的光束发射。在旋转盘体模的仿真中,观察到角度估计具有360°均匀行为,中位角度偏差为1.01°,中位角度SD为1.8°。在直线容器上进行模拟和测量时都获得了相似的结果,其中角度偏差低于1.5°,SD约为1°。在仿真和测量中,估计的速度大小也保持在10%偏差和5%相对SD下。对健康个体的颈动脉分叉进行体内测量。捕获三个心动周期中的3秒采集。在心脏收缩期,在颈动脉中观察到一致且重复的涡流。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号