首页> 外文期刊>Ultrasonics, Ferroelectrics and Frequency Control, IEEE Transactions on >A new method for blood velocity measurements using ultrasound FMCW signals
【24h】

A new method for blood velocity measurements using ultrasound FMCW signals

机译:使用FMCW超声波信号进行血流速度测量的新方法

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The low peak power of frequency-modulated continuous wave (FMCW) radar makes it attractive for various applications, including vehicle collision warning systems and airborne radio altimeters. This paper describes a new ultrasound Doppler measurement system that measures blood flow velocity based on principles similar to those of FMCW radar. We propose a sinusoidal wave for FM modulation and introduce a new demodulation technique for obtaining Doppler information with high SNR and range resolution. Doppler signals are demodulated with a reference FMCW signal to adjust delay times so that they are equal to propagation times between the transmitter and the receiver. Analytical results suggest that Doppler signals can be obtained from a selected position, as with a sample volume in pulse wave Doppler systems, and that the resulting SNR is nearly identical to that obtained with continuous wave (CW) Doppler systems. Additionally, clutter power is less than that of CW Doppler systems. The analytical results were verified by experiments involving electronic circuits and Doppler ultrasound phantoms.
机译:调频连续波(FMCW)雷达的低峰值功率使其对各种应用具有吸引力,包括车辆碰撞预警系统和机载无线电高度计。本文介绍了一种新的超声多普勒测量系统,该系统基于类似于FMCW雷达的原理测量血流速度。我们提出了用于FM调制的正弦波,并介绍了一种新的解调技术,以获得具有高SNR和距离分辨率的多普勒信息。用参考FMCW信号对多普勒信号进行解调,以调整延迟时间,以使它们等于发射器和接收器之间的传播时间。分析结果表明,可以从选定的位置获得多普勒信号,就像脉搏多普勒系统中的样本量一样,并且所得的SNR几乎与连续波(CW)多普勒系统获得的SNR相同。另外,杂波功率小于连续波多普勒系统。通过涉及电子电路和多普勒超声体模的实验验证了分析结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号