首页> 外文期刊>IEEE Transactions on Nuclear Science >High rate large dynamic range analog circuitry and digitizers for fast calorimetry
【24h】

High rate large dynamic range analog circuitry and digitizers for fast calorimetry

机译:高速率大动态范围模拟电路和数字化仪,用于快速量热

获取原文
获取原文并翻译 | 示例

摘要

This paper describes the issues encountered in the design of a large dynamic range digitizer for calorimetry. A gain switching technology developed for the liquid krypton calorimeter of the NA48 CP-violation experiment at CERN is described in detail. 14 bit dynamic range and better than 300 ps time resolution were achieved in beam tests with an 8 channel prototype. A mixed analog/digital ASIC has been developed. A 13,500 channel system is being produced which will be put into operation in 1996.
机译:本文介绍了用于量热法的大型动态范围数字化仪设计中遇到的问题。详细介绍了为CERN的NA48 CP违反实验的液体k量热仪开发的增益切换技术。在使用8通道原型的波束测试中,实现了14位动态范围和优于300 ps的时间分辨率。已经开发了混合的模拟/数字ASIC。正在生产13,500个频道系统,该系统将于1996年投入运行。