The choice of pinhole geometry is a critical factor in the performance of pinhole-collimator-based source tracking systems for brachytherapy QA. In this work, an analytical model describing the penetrative sensitivity of a single-cone pinhole collimator to photons emitted from a point source is derived. Using existing models for single-cone resolution and double-cone sensitivity and resolution, the theoretical sensitivity and resolution of the single-cone collimator are quantitatively compared with those of a double-cone collimator with an equivalent field of view. Monte Carlo simulations of the single and double-cone pinhole collimators using an accurate 3D model of a commercial high dose rate brachytherapy source are performed to evaluate the relative performance of each geometry for a novel real-time HDR brachytherapy QA system, HDR BrachyView. The theoretical penetrative sensitivity of the single-cone pinhole is shown to be higher than the double-cone pinhole, which is in agreement with the results from the Monte Carlo simulations. The wider pinhole response function of the single-cone collimator results in a larger total error between the projected center of the source and the estimated center of mass of the source projection for the single-cone collimator, with the greatest error (at the maximum FoV angle) being 0.54 mm for the double-cone pinhole and 1.37 mm for the single-cone at . The double-cone pinhole geometry is determined to be the most appropriate choice for the pinhole collimator in the HDR BrachyView probe.
机译:针孔几何形状的选择是用于近距离放射治疗QA的基于针孔准直器的源跟踪系统性能的关键因素。在这项工作中,得出了一个分析模型,该模型描述了单锥孔针状准直仪对从点源发出的光子的穿透敏感性。使用现有的单锥度准直器和双锥度准直器的模型,将具有等效视场的双锥度准直器的理论灵敏度和分辨率定量比较。使用商业高剂量率近距离放射治疗源的精确3D模型对单锥孔针孔准直器和双锥形针孔准直仪进行蒙特卡罗模拟,以评估新型实时HDR近距离放射治疗QA系统HDR BrachyView的每种几何形状的相对性能。单锥孔的理论穿透敏感性高于双锥孔,这与蒙特卡洛模拟的结果一致。单锥形准直器的针孔响应函数更宽,导致单锥形准直器的源投影中心与源投影的估计质心之间的总误差更大,且误差最大(在最大FoV下)双锥针孔的最大角度为0.54 mm,单锥针孔的最大角度为1.37 mm。对于HDR BrachyView探针中的针孔准直仪,确定双锥针孔几何形状是最合适的选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号