Brachyview: An in-body imaging system for real-time QA in HDR prostate brachytherapy

机译:Brachyview:用于HDR前列腺近距离放射治疗的实时QA的体内成像系统

获取原文
获取原文并翻译 | 示例

摘要

A transrectal real-time source tracking system, BrachyView, is developed by Centre for Medical Radiation Physics at University of Wollongong for treatment quality control during the high dose rate prostate brachytherapy procedure (HDR-PBT). The probe consists of a 15 × 60 mm2 pixellated silicon detector (four Timepix detectors) encased in a cylindrical seven pinhole tungsten collimator. The source position is determined by back-projecting the source projection images through the collimator pinholes. A prototype BrachyView probe consisting of a single Timepix detector (15 × 15 mm2) and a planar seven pinhole collimator is evaluated in this paper. The source position can be estimated with a maximum error of 0.2 mm on the detector plane, which translates into a maximum error of 1.42 mm within the prostate volume. This result is in good agreement with the results obtained in previously published simulation studies.
机译:卧龙岗大学医学放射物理中心开发了一种经直肠实时源跟踪系统BrachyView,用于在高剂量率前列腺近距离放射治疗程序(HDR-PBT)期间控制治疗质量。该探头由一个15×60 mm 2 像素化硅探测器(四个Timepix探测器)组成,该探测器装在圆柱形的七个针孔钨准直仪中。通过通过准直仪针孔反投影源投影图像来确定源位置。本文评估了原型BrachyView探针,该探针由单个Timepix检测器(15×15 mm 2 )和一个平面的七个针孔准直仪组成。可以在检测器平面上以0.2毫米的最大误差估算源位置,这将转化为前列腺体积内的1.42毫米的最大误差。该结果与先前发表的模拟研究中获得的结果高度吻合。