Robust Sensing of Distal Pulse Waveforms on a Modified Weighing Scale for Ubiquitous Pulse Transit Time Measurement

机译:用于无处不在的脉冲传输时间测量的改进型秤上的远距离脉冲波形的鲁棒性检测

获取原文
获取原文并翻译 | 示例
           

摘要

The measurement of aortic pulse transit time (PTT), the time for the arterial pulse wave to travel from the carotid to the femoral artery, can provide valuable insight into cardiovascular health, specifically regarding arterial stiffness and blood pressure (BP). To measure aortic PTT, both proximal and distal arterial pulse timings are required. Recently, our group has demonstrated that the ballistocardiogram signal measured on a modified weighing scale can provide an unobtrusive, yet accurate, means of obtaining a proximal timing reference; however, there are no convenient, reliable methods to extract the distal timing from a subject standing on the modified weighing scale. It is common to use a photoplethysmograph (PPG) attached to a toe to measure this distal pulse, but we discovered that this signal is greatly deteriorated as the subject stands on the scale. In this paper, we propose a novel method to measure the distal pulse using a custom reflective PPG array attached to the dorsum side of the foot (D-PPG). A total of 12 subjects of varying skin tones were recruited to assess the preliminary validation of this approach. Pulse measurements using the D-PPG were taken from seated and standing subjects, and the commercially available PPG were measured for facilitating comparison of timing measurements. We show that the D-PPG was the only sensor to retain the high detection rate of feasible timing values. To further test and optimize the system, various factors such as applied pressure, measurement location, and LED/photodiode configuration were tested.
机译:测量主动脉搏动传播时间(PTT),即动脉搏动波从颈动脉行至股动脉的时间,可以为心血管健康提供有价值的见解,尤其是在动脉僵硬度和血压(BP)方面。为了测量主动脉PTT,需要近端和远端动脉脉冲定时。最近,我们的小组已经证明,在修改后的体重秤上测量的心电图信号可以提供一种简便易行但又准确的方法来获取近端计时参考。但是,没有方便,可靠的方法从站立在修改后的体重秤上的受试者中提取远端时间。通常使用附在脚趾上的光电容积描记器(PPG)来测量该远端脉搏,但是我们发现,随着受试者站在体重秤上,该信号会大大恶化。在本文中,我们提出了一种新方法,该方法使用附着在脚背侧(D-PPG)上的定制反射式PPG阵列来测量远端脉冲。总共招募了12名不同肤色的受试者,以评估这种方法的初步验证。使用D-PPG进行的脉搏测量来自就座和站立的受试者,并测量了商用PPG以便于比较计时测量。我们表明D-PPG是唯一保持可行定时值的高检测率的传感器。为了进一步测试和优化系统,测试了各种因素,例如施加压力,测量位置和LED /光电二极管配置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号