Differences in mean arterial pressure of young and elderly people measured by oscilometry during inflation and deflation of the arm cuff

机译:臂袖充气和放气期间通过振动法测量的年轻人和老年人的平均动脉压差异

获取原文
获取原文并翻译 | 示例
           

摘要

Systemic arterial blood pressure (BP) is one of the most important parameters of the cardiovascular system. An oscillometric NIBP monitor was specifically designed to measure oscillometric pulsations and mean arterial pressure (MAP) during inflation and deflation of the cuff. Nineteen healthy young (age 23.1 +/- 1.7 years; mean +/- SD) and 35 elderly (83.9 +/- 7.9 years; mean +/- SD) subjects were studied. Differential analysis of MAP during inflation and deflation show mean |Delta MAP| = 2.9 +/- 2.6 mm Hg in the young group (mean +/- SD) and vertical bar Delta MAP| = 6.3 +/- 5.2 mm Hg for seniors (mean +/- SD). There was a significant difference (p < 0.05) in means of |Delta MAP| measured during cuff inflation and cuff deflation between both groups. In about 50% of elderly subjects |Delta MAP| was higher than 5 mm Hg. Potential clinical relevance of the method needs to be further evaluated.
机译:全身动脉血压(BP)是心血管系统最重要的参数之一。示波NIBP监护仪专门用于测量袖带充气和放气期间的示波脉动和平均动脉压(MAP)。研究对象为19名健康的年轻人(年龄23.1 +/- 1.7岁;平均+/- SD)和35名老年人(83.9 +/- 7.9岁;平均+/- SD)。充气和放气期间MAP的差异分析显示平均值| Delta MAP | =年轻人组中2.9 +/- 2.6毫米汞柱(平均值+/- SD)和垂直条Delta =老年人6.3 +/- 5.2毫米汞柱(平均+/- SD)。 | Delta MAP |的均值有显着差异(p <0.05)。在两组之间的袖带充气和袖带放气期间进行测量。在大约50%的老年受试者中| Delta MAP |高于5毫米汞柱。该方法潜在的临床相关性需要进一步评估。

著录项

  • 来源
    《Biomedizinische Technik》 |2016年第6期|611-621|共11页
  • 作者单位

    Czech Tech Univ, Fac Elect Engn, Dept Phys, Prague, Czech Republic;

    Czech Tech Univ, Fac Elect Engn, Dept Circuit Theory, Prague, Czech Republic;

    Czech Tech Univ, Fac Elect Engn, Dept Circuit Theory, Prague, Czech Republic;

    Czech Tech Univ, Czech Inst Informat Robot & Cybernet, Prague, Czech Republic;

    St Annes Univ, Fac Hosp, Int Clin Res Ctr, Dept Internal Med Cardioangiol, Brno, Czech Republic|Mayo Clin, Dept Internal Med, Div Cardiovasc Dis, Rochester, NY USA;

    Czech Tech Univ, Fac Elect Engn, Dept Circuit Theory, Prague, Czech Republic;

    Mayo Clin, Dept Internal Med, Div Cardiovasc Dis, Rochester, NY USA;

    Czech Tech Univ, Fac Elect Engn, Dept Circuit Theory, Prague, Czech Republic;

    Mayo Clin, Dept Internal Med, Div Cardiovasc Dis, Rochester, NY USA;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    blood pressure measurement; mean arterial pressure; oscillometric method;

    机译:血压测量;平均动脉压;示波法;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号