首页> 外文期刊>European Radiology >Characterization of microcalcification: can digital monitor zooming replace magnification mammography in full-field digital mammography?
【24h】

Characterization of microcalcification: can digital monitor zooming replace magnification mammography in full-field digital mammography?

机译:微钙化的特征:数字监视器缩放可以代替全场数字乳腺摄影中的放大乳腺摄影吗?

获取原文
获取原文并翻译 | 示例

摘要

The aim of this study was to compare the diagnostic accuracy and image quality of microcalcifications in zoomed digital contact mammography with digital magnification mammography. Three radiologists with different levels of experience in mammography reviewed 120 microcalcification clusters in 111 patients with a full-field digital mammography system relying on digital magnification mammogram (MAG) images and zoomed images from contact mammography (ZOOM) using commercially available zooming systems on monitors. Each radiologist estimated the probability of malignancy and rated the image quality and confidence rate. Performance was evaluated by sensitivity, specificity, positive predictive value, negative predictive value, and receiver operating characteristic (ROC) analysis. All three radiologists rated MAG images higher than ZOOM images for sensitivity with statistical significance (average value, 92% vs. 87%, P < 0.05) and performance by ROC analysis improved with MAG imaging. The confidence rate for diagnosis decision and the assessment of lesion characteristics were also better in MAG images than in ZOOM images with statistical significance (P < 0.0001). Digital magnification mammography can enhance diagnostic performance when characterizing microcalcifications. Images zoomed from digital contact mammography cannot serve as an alternative to direct magnification digital mammography.
机译:这项研究的目的是比较放大数字接触式乳腺X线摄影术与数字放大乳腺X线摄影术中微钙化的诊断准确性和图像质量。三位在乳腺X射线照相方面经验不同的放射科医生回顾了111例患者的120个微钙化簇,这些患者依靠全数字乳腺X线照片(MAG)图像和接触式乳腺X射线照片(ZOOM)的放大图像,并使用监视器上的市售变焦系统。每位放射科医生估计恶性肿瘤的可能性,并评估图像质量和置信度。通过敏感性,特异性,阳性预测值,阴性预测值和受试者工作特征(ROC)分析来评估表现。三位放射科医生对MAG图像的敏感性均高于ZOOM图像,具有统计学意义(平均值,92%对87%,P <0.05),并且通过ROM成像提高了ROC分析的性能。 MAG图像的诊断决策和病变特征评估的置信度也比具有统计学意义的ZOOM图像更好(P <0.0001)。当表征微钙化时,数字放大乳房X线照相术可以增强诊断性能。从数字接触式乳腺摄影术放大的图像不能代替直接放大数字乳腺摄影术。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号