Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT

机译:冠状动脉支架多探测器冠状动脉CT血管造影的最新进展:双源CT对29种不同支架类型的体外评估

获取原文
获取原文并翻译 | 示例

摘要

The aim of this study was to test a large sample of the latest coronary artery stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation, and image noise in dual-source multidetector row CT (DSCT) in vitro and to provide a CT catalogue of currently used coronary artery stents. Twenty-nine different coronary artery stents (19 steel, 6 cobalt-chromium, 2 tantalum, 1 iron, 1 magnesium) were examined in a coronary artery phantom (vessel diameter 3 mm, intravascular attenuation 250 HU, extravascular density −70 HU). Stents were imaged in axial orientation with standard parameters: 32 × 0.6 collimation, pitch 0.24, 400 mAs, 120 kV, rotation time 0.33 s. Image reconstructions were obtained with four different convolution kernels (soft, medium-soft, standard high-resolution, stent-dedicated). To evaluate visualization characteristics of the stent, the lumen diameter, intraluminal density, and noise were measured. The stent-dedicated kernel offered best average lumen visualization (54 ± 8.3%) and most realistic lumen attenuation (222 ± 44 HU) at the expense of increased noise (23.9 ± 1.9 HU) compared with standard CTA protocols (p < 0.001 for all). The magnesium stent showed the least artifacts with a lumen visibility of 90%. The majority of stents (79%) exhibited a lumen visibility of 50–59%. Less than half of the stent lumen was visible in only six stents. Stent lumen visibility largely varies depending on the stent type. Magnesium is by far more favorable a stent material with regard to CT imaging when compared with the more common materials steel, cobalt-chromium, or tantalum. The magnesium stent exhibits a lumen visibility of 90%, whereas the majority of the other stents exhibit a lumen visibility of 50–59%.
机译:这项研究的目的是在体外双源多探测器行CT(DSCT)中使用四种图像重建方法在管腔可视化,管腔衰减和图像噪声方面测试大量最新的冠状动脉支架样品,并提供一种当前使用的冠状动脉支架的CT目录。在冠状动脉模型中检查了29个不同的冠状动脉支架(19个钢,6个钴铬,2个钽,1个铁,1个镁)(血管直径3 mm,血管内衰减250 HU,血管外密度-70 HU)。使用标准参数:32×0.6准直,间距0.24、400 mAs,120 kV,旋转时间0.33 s的轴向取向的支架成像。使用四个不同的卷积核(软,中软,标准高分辨率,支架专用)获得图像重建。为了评估支架的可视化特性,测量了管腔直径,管腔内密度和噪声。与标准CTA方案相比,专用于支架的内核提供了最佳的平均流明可视性(54±8.3%)和最真实的流明衰减(222±44 HU),但以增加的噪声(23.9±1.9 HU)为代价(所有P值均<0.001) )。镁支架的伪影最少,管腔可见度为90%。大多数支架(79%)的内腔可见度为50-59%。仅六个支架中可见不到支架内腔的一半。支架内腔的可见性在很大程度上取决于支架的类型。与更常见的钢,钴铬或钽材料相比,镁在CT成像方面是更有利的支架材料。镁支架的内腔可见度为90%,而其他大多数支架的内腔可见度为50–59%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号