首页> 美国卫生研究院文献>Radiology Research and Practice >Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types
【2h】

Dual-Source CT Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types

机译:外周动脉支架的双源CT血管造影:22种不同类型支架的体外评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose. To test different peripheral arterial stents using four image reconstruction approaches with respect to lumen visualization, lumen attenuation and image noise in dual-source multidetector row CT (DSCT) in vitro. Methods and Materials. 22 stents (nitinol, steel, cobalt-alloy, tantalum, platinum alloy) were examined in a vessel phantom. All stents were imaged in axial orientation with standard parameters. Image reconstructions were obtained with four different convolution kernels. To evaluate visualization characteristics of the stent, the lumen diameter, intraluminal density and noise were measured. Results. The mean percentage of the visible stent lumen diameter from the nominal stent diameter was 74.5% ± 5.7 for the medium-sharp kernel, 72.8% ± 6.4 for the medium, 70.8% ± 6.4 for the medium-smooth and 67.6% ± 6.6 for the smooth kernel. Mean values of lumen attenuation were 299.7HU ± 127 (medium-sharp), 273.9HU ± 68 (medium), 270.7HU ± 53 (medium-smooth) and 265.8HU ± 43. Mean image noise was: 54.6 ± 6.3, 20.5 ± 1.7, 16.3 ± 1.7, 14.0 ± 2 respectively. Conclusion. Visible stent lumen diameter varies depending on stent type and scan parameters. Lumen diameter visibility increases with the sharpness of the reconstruction kernel. Smoother kernels provide more realistic density measurements inside the stent lumen and less image noise.
机译:目的。为了在管腔可视化,管腔衰减和图像噪声方面采用四种图像重建方法在体外双源多探测器行CT(DSCT)中测试不同的外周动脉支架。方法和材料。在血管模型中检查了22个支架(镍钛诺,钢,钴合金,钽,铂合金)。用标准参数在轴向取向上对所有支架成像。使用四个不同的卷积核获得了图像重建。为了评估支架的可视化特性,测量了管腔直径,管腔内密度和噪声。结果。中尖锐内核的可见支架内腔直径相对于标称支架直径的平均百分比为:中等尖粒仁的为74.5%±5.7,中等光滑粒为72.8%±6.4,中等光滑粒为70.8%±6.4,中等光滑粒为67.6%±6.6。光滑的内核。流明衰减平均值分别为299.7HU±127(中等锐度),273.9HU±68(中等),270.7HU±53(中等平滑度)和265.8HU±43.平均图像噪声分别为:54.6±6.3、20.5±0.5分别为1.7、16.3±1.7、14.0±2。结论。可见的支架管腔直径根据支架类型和扫描参数而异。流明直径的可见度随重建核的清晰度而增加。较光滑的内核可在支架内腔内部提供更真实的密度测量值,并减少图像噪声。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号