MR Angiography of Peripheral Arterial Stents: In Vitro Evaluation of 22 Different Stent Types

机译:外周动脉支架的MR血管造影:22种不同类型支架的体外评估

获取原文
       

摘要

Purpose. To evaluate stent lumen visibility of a large sample of different peripheral arterial (iliac, renal, carotid) stents using magnetic resonance angiography in vitro.Materials and Methods. 21 different stents and one stentgraft (10 nitinol, 7 316L, 2 tantalum, 1 cobalt superalloy, 1 PET + cobalt superalloy, and 1 platinum alloy) were examined in a vessel phantom (vessel diameters ranging from 5 to 13 mm) filled with a solution of Gd-DTPA. Stents were imaged at 1.5 Tesla using a T1-weighted 3D spoiled gradient-echo sequence. Image analysis was performed measuring three categories: Signal intensity in the stent lumen, lumen visibility of the stented lumen, and homogeneity of the stented lumen. The results were classified using a 3-point scale (good, intermediate, and poor results).Results. 7 stents showed good MR lumen visibility (4x nitinol, 2x tantalum, and 1x cobalt superalloy). 9 stents showed intermediate results (5x nitinol, 2x 316L, 1x PET + cobalt superalloy, and 1x platinum alloy) and 6 stents showed poor results (1x nitinol, and 5x 316L).Conclusion. Stent lumen visibility varies depending on the stent material and type. Some products show good lumen visibility which may allow the detection of stenoses inside the lumen, while other products cause artifacts which prevent reliable evaluation of the stent lumen with this technique.
机译:目的。使用体外磁共振血管成像评估一大批不同外周动脉((,肾,颈动脉)支架的支架内腔可见性。材料和方法。在一个装满血管的体模中(血管直径为5到13毫米),检查了21种不同的支架和一种支架移植物(10根镍钛合金,7 316L,2钽,1钴高温合金,1 PET +钴高温合金和1铂合金)。 Gd-DTPA解决方案。使用T1加权3D渐变梯度回波序列在1.5 Tesla下对支架成像。进行了图像分析,测量了以下三类:支架内腔中的信号强度,支架内腔的内腔可见性以及支架内腔的均匀性。使用3分制对结果进行分类(好,中和差)。 7个支架显示出良好的MR管腔可见性(4个镍钛合金,2个钽合金和1个钴高温合金)。 9个支架显示中间结果(5x镍钛合金,2x 316L,1x PET +钴超合金和1x铂合金),6个支架显示较差结果(1x镍钛合金和5x 316L)。支架内腔的可见性取决于支架的材料和类型。一些产品显示出良好的管腔可见性,这可能允许检测管腔内部的狭窄,而其他产品会导致伪影,从而妨碍使用该技术可靠评估支架管腔。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号