首页> 外文期刊>European Journal of Applied Physiology >Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus
【24h】

Evaluation of three portable blood lactate analysers: Lactate Pro, Lactate Scout and Lactate Plus

机译:评估三种便携式血液乳酸分析仪:Lactate Pro,Lactate Scout和Lactate Plus

获取原文
获取原文并翻译 | 示例
           

摘要

Three portable blood lactate analysers, Lactate Pro (LP), Lactate Scout (LS) and Lactate Plus (L+), were evaluated. Analyser reliability and accuracy was assessed. For reliability, intra- and inter-analyser comparisons demonstrated that the LP (intra-TE = 0.5 mM, inter-TE = 0.4 mM) and L+ (intra-TE = 0.4, inter-TE = 0.4 mM) displayed greater overall reliability than the LS (intra-TE = 1.0, inter-TE = 0.8 mM). At BLa 8.0 mM, the LP (intra-TE = 0.5 mM, inter-TE = 0.4 mM) and L+ (intra- and inter-TE = 0.4 mM) displayed greater reliability than the LS (intra-TE = 1.1 mM, inter-TE = 0.9 mM). For accuracy, the L+ (SEE = 0.6 mM) compared more favourably to the LP than the LS (SEE = 1.1 mM). At BLa ~1.0–18.0 mM, the LS produced values that were up to 0.9 mM higher than the LP; the L+ produced BLa that were within ±0.1 mM. All portable analysers tended to under-read the ABL 700 analyser. The suitability of the LP and L+ as accurate analysers is supported by strong correlations (r = 0.91 and r = 0.94) and limits of agreement ≤2.1 mM. This study showed that the LP and L+, compared well to each other, displayed good reliability and accuracy when compared to a laboratory-based analyser. Although the LS also displayed relatively good reliability, it was not as reliable or accurate as the LP or L+.
机译:评估了三种便携式血液乳酸分析仪:Lactate Pro(LP),Lactate Scout(LS)和Lactate Plus(L + )。分析仪的可靠性和准确性进行了评估。为了可靠性,分析器内和分析器间的比较表明,LP(TE内= 0.5 mM,TE间= 0.4 mM)和L + (TE内= 0.4,TE间= 0.4 mM)显示的整体可靠性高于LS(TE内= 1.0,TE间= 0.8 mM)。在BLa 8.0 mM时,LP(TE内= 0.5 mM,TE间= 0.4 mM)和L + (TE内和TE = 0.4 mM)显示出比LS( TE内= 1.1 mM,TE内= 0.9 mM)。出于准确性考虑,L + (SEE = 0.6 mM)比LP比LS(SEE = 1.1 mM)更适合LP。在BLa〜1.0-18.0 mM时,LS产生的值比LP高出0.9 mM。 L + 产生的BLa在±0.1 mM之内。所有便携式分析仪都倾向于读取ABL 700分析仪。 LP和L + 作为精确分析仪的适用性受到强大的相关性(r = 0.91和r = 0.94)和一致性限制≤2.1mM的支持。这项研究表明,与基于实验室的分析仪相比,LP和L + 相互比较好,显示出良好的可靠性和准确性。尽管LS也显示出相对较好的可靠性,但它不如LP或L + 可靠或准确。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号