首页> 外文期刊>Australian and New Zealand Journal of Obstetrics and Gynecology >Lactate study using umbilical cord blood: Agreement between Lactate Pro hand-held devices with blood gas analyser and evaluation of lactate stability over time
【24h】

Lactate study using umbilical cord blood: Agreement between Lactate Pro hand-held devices with blood gas analyser and evaluation of lactate stability over time

机译:使用脐带血进行乳酸研究:带有血气分析仪的Lactate Pro手持设备之间的协议以及随着时间的推移评估乳酸稳定性

获取原文
获取原文并翻译 | 示例
           

摘要

Background Lactate measurements have become increasingly preferred over pH analysis in the evaluation of fetal acidaemia in labour. In a busy labour ward, often the umbilical cord may be sampled late and as a result yield unreliable lactate values. Aim To investigate the agreement of hand-held device Lactate Pro with a reference method blood gas analyser and evaluate the stability of umbilical cord lactate values over time. Methods Prospective study carried out at elective caesarean section. Sixteen umbilical cords were double clamped immediately after delivery with paired arterial and venous blood samples collected by an independent researcher, at varying time intervals, and processed by two Lactate Pro devices and a reference method blood gas analyser. Results A significant difference of -0.41 to 0.10 mmol/L was found when different groups of Lactate Pro devices were compared with blood gas analyser at lactate values up to 5.70 mmol/L, with average lactate value of 2.45 mmol/L. Over time, there is progressive rise in lactate samples obtained from the umbilical cord. Conclusion Lactate Pro devices have a significant difference, but when used in clinical practice on cord blood after delivery, this is unlikely to be meaningful. In intrapartum fetal surveillance, a systematic overestimation might lead to unnecessary intervention. It is possible to retrospectively predict the likely level of lactate at birth in delayed cord samples.
机译:背景技术在劳动中胎儿酸血症的评估中,乳酸测量已变得比pH分析越来越受欢迎。在繁忙的劳动病房中,脐带常常采样较晚,结果产生的乳酸值不可靠。目的研究手持设备Lactate Pro与参考方法血气分析仪的一致性,并评估随时间推移脐带乳酸值的稳定性。方法对选择性剖宫产进行前瞻性研究。分娩后立即将16条脐带与独立研究人员在不同时间间隔收集的成对的动脉和静脉血样本双夹,并通过两个Lactate Pro设备和参考方法血气分析仪进行处理。结果当不同组的Lactate Pro设备与血气分析仪进行比较时,乳酸盐值高达5.70 mmol / L,平均乳酸盐值是2.45 mmol / L,发现-0.41至0.10 mmol / L有显着差异。随着时间的流逝,从脐带获得的乳酸盐样品逐渐增多。结论Lactate Pro设备具有显着差异,但是在分娩后用于脐带血的临床实践中时,这不太可能有意义。在分娩期胎儿监护中,系统的高估可能导致不必要的干预。可以回顾性地预测延迟脐带样本出生时乳酸的水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号