In the electromechanical type indirect pressure indicating devices (IPIDs), the applied pressure is indirectly computed by some relationship between applied pressure and an analog / digital electrical signal in terms of voltage, current or frequency measured by a sensing element of such devices. If the output of an EMT in terms of any electrical quantity i.e. voltage, current or frequency, is related / dependent upon input quantity i.e. pressure then this relationship can be defined by a curve that has been fitted to experimental data by method of least square. The calculation of the calibration results and the associated uncertainties is a complex matter involving many influencing quantities. The present paper describes a method for calibration and evaluating uncertainty in pressure measurement using such EMTs with EOs through statistical analysis of errors.
机译:在机电类型的间接压力指示装置(IPID)中,通过施加压力与模拟/数字电信号之间的某种关系来间接计算施加压力,该关系取决于由此类装置的感测元件测量的电压,电流或频率。如果以任何电量(即电压,电流或频率)表示的EMT的输出与输入量(即压力)相关/取决于输入量(即压力),则可以通过一条曲线来定义这种关系,该曲线已通过最小二乘法拟合到实验数据。校准结果和相关不确定性的计算是一个涉及许多影响因素的复杂问题。本文介绍了一种通过对误差进行统计分析,使用带有EO的EMT来校准和评估压力测量不确定度的方法。