Evaluation of point-of-care testing device for anemia detection: A cross-sectional method comparison study from Thailand

机译:评估贫血检测点检测装置:泰国截面法对比较研究

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background A comparison study is crucial before launching a new medical device; therefore, we compared the Mission Ultra Hb Testing System with the Sysmex XN-3000 automated hematology analyzer in Thai adult males and non-pregnant adult females. Methods Parallel studies were conducted using discarded venous K2-ethylenediaminetetraacetic acid samples from participants requiring hematological investigations. According to the World Health Organization criteria, the participants were categorized as overall, anemia, and non-anemia for analysis. Results Three hundred participants were included in this study. In all participants, near-perfect correlation and agreement were observed between the two methods for Hb measurement ( r =?0.963, p ?0.001) with an interclass correlation coefficient (ICC) of 0.981 (95% confidence interval [CI]: 0.976–0.985) and Hct measurement ( r =?0.941, p ?0.001) with an ICC of 0.965 (95% CI: 0.956–0.972). The sensitivity and specificity of the device in detecting anemia were 86.2% (95% CI: 79.7–91.2) and 98.6% (95% CI: 95.2–99.8), respectively. The area under the curve was 0.976 (95% CI: 0.963–0.989). The device showed average biases of 0.76?g/dl (95% limits of agreement [LOA]: ?1.03 to 2.54) for Hb measurement and ?2.73% (95% LOA: ?9.28 to 3.82) for Hct measurement in all participants. Conclusion Agreement between the Mission Ultra Hb Testing System and Sysmex XN-3000 was observed. The device was excellent for detecting anemia. However, the essential evidence showing biases of the Hb and Hct measurements obtained from the device was revealed. Laboratory interpretation should be carefully performed, particularly at the near cut-off values.
机译:背景技术比较研究在发射新医疗设备之前至关重要;因此,我们将特派团超HB检测系统与Sysmex XN-3000自动血液分析仪进行了比较了泰国成年男性和非怀孕成年女性。方法使用来自需要血液学调查的参与者的丢弃的静脉K2-乙二胺四乙酸样品进行并行研究。根据世界卫生组织标准,参与者分为整体,贫血和非贫血症分析。结果本研究纳入了三百名参与者。在所有参与者中,在HB测量的两种方法之间观察到近乎完美的相关性和协议(R = 0.963,P& 0.001),其中杂类相关系数(ICC)为0.981(95%置信区间[CI]: 0.976-0.985)和HCT测量(R = 0.941,P& 0.001),ICC为0.965(95%CI:0.956-0.972)。检测贫血中的装置的敏感性和特异性分别为86.2%(95%CI:79.7-91.2)和98.6%(95%CI:95.2-99.8)。曲线下的区域为0.976(95%CI:0.963-0.989)。该装置显示出0.76?G / DL的平均偏差(协议的95%限制:?1.03至2.54)用于HB测量,并且?2.73%(95%LOA:?9.28至3.82),用于所有参与者的HCT测量。结论特派团在特派团超HB检测系统和Sysmex XN-3000之间的协议。该装置非常适合检测贫血。然而,揭示了显示HB的偏差和从装置获得的HCT测量的基本证据。应仔细执行实验室解释,特别是在接近截止值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号