Evaluation of the Structural Behaviour of a Unilateral External Fixator for Osteosynthesis

机译:对骨质合成的单侧外固定器的结构行为评价

获取原文

摘要

Background:External fixation is an osteosynthesis technique particularly useful in trauma surgery and Damage Control Orthopedics (DCO). However, complications, such as pin loosening and pin tract infections, are fairly common. For reducing thermal damage and infection rates, monocortical pins have been proposed as an alternative to the most used bicortical pins. However, there is a lack of studies regarding their mechanical properties.Objective:The aim of the study is to assess the static and dynamic stability of a unilateral external fixator experimentally when applied through monocortical pins for the reduction of femur and tibia fractures.Methods:A modular unilateral external fixator was used and a total number of 6 pins were used per test. The static tests were performed in displacement control by applying a vertical displacement to the upper fixture at 1 mm/min until a tension load of 380 N was reached. The dynamic tests were performed by applying a sinusoidal displacement. During each test, forces and crosshead displacements were acquired. Two different stiffness indexes were assessed.Results: By comparing the two anatomic regions, it was found that the fixator behaves stiffer when mounted on the femur, regardless of the pins used, while stiffnesses comparable to the femur ones are reached by the tibia when 4 mm diameter pins are used. Static analysis revealed excellent fixator stability when implanted with 4 mm diameter monocortical pins on both anatomic regions. On the contrary, two tibia and one femur samples showed failures at the bone-pin interface when 3 mm diameter pins were used.Conclusion:Dynamic analysis showed no substantial difference between the tested configurations and confirmed the fixator's ability to sustain cyclic loading without further damage to the sample.
机译:背景:外部固定是一种骨质合成技术,特别适用于创伤手术和损伤控制骨科(DCO)。然而,并发症,例如别针松动和别针感染,相当普遍。为了减少热损伤和感染率,已提出单根销作为最常用的双模销的替代品。然而,缺乏关于其机械性质的研究。目的是在通过单眼销施加的股骨和胫骨骨折的单眼销施用时通过实验评估单侧外固定器的静态和动态稳定性。方法:使用模块化单侧外部固定器,每次测试使用总数为6个引脚。通过将垂直位移施加到1mm / min的上固定装置,直到达到380n的张力载荷,在位移控制中进行静态试验。通过施加正弦位移来进行动态测试。在每个测试期间,获得力和交叉头位移。评估两种不同的刚度指数。结果:通过比较两个解剖区域,发现固定器在安装在股骨上时更硬,而不管使用的销钉,胫骨相当的刚度达到胫骨时达到胫骨使用直径销。静态分析显示出在两个解剖区域上植入4毫米直径的单根引脚时出色的固定器稳定性。相反,当使用3毫米的销时,两种胫骨和一个股骨样品在骨引脚界面处显示出故障。结论:动态分析在测试的配置之间没有显着差异,并确认了固定器维持循环载荷的能力而不会进一步损坏到样本。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号