Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading

机译:机械载荷施加后四种不同校长螺钉扭矩对单植入式假体螺杆松动的影响

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Background . The complications of implant-supported prostheses can be classified into mechanical and biological ones, one part of which is associated with screw loosening. This study was aimed to compare the effect of four different abutment screw torque techniques on screw loosening in single implant-supported prostheses following the application of mechanical loading. Materials and Methods . In this experimental study, a total of 40 implants in acrylic blocks (6?×?10?×?20?mm) were mounted perpendicular to the surface. They were then randomly divided into four groups: (1) torquing once with 30?Ncm, (2) torquing three times with 30?Ncm and 5-minute intervals, (3) torquing once with 30?Ncm, opening the screw, and retorquing with 30?Ncm, and (4) torquing once with 35?Ncm. The torque values were confirmed by using a digital torque meter. Then, the samples underwent a force (2?cps, 0.453–11.793?kg) for three hours before the measurement of detorque values. The screw loosening force (torque) was then measured and recorded. The obtained data were analyzed by SPSS (version 22) software using one-way ANOVA and Tukey post hoc test at a 5% error level. Results . The maximum mean detorque values of the abutment screws in single implant-supported prostheses were reported for groups 4 (27.8?±?1.3), 1 (26.8?±?1.3), and 3 (25.1?±?1.3), and the minimum mean detorque value was found in group 2 (24.9?±?1.2). Moreover, no significant difference was observed between groups 2 and 3 ( ), but a significant difference was found between groups 1 and 3 and other groups ( ). Conclusion . The increase in the torque value increased the torque loss. However, the detorque value in group 4 showed the least difference with the value recommended by the manufacturer (30?Ncm).
机译:背景 。植入物支持的假体的并发症可以分为机械和生物学,其中一部分与螺杆松动有关。本研究旨在比较在施加机械载荷后单个植入物支持的假体中螺杆松开的螺杆松动的影响。材料和方法 。在该实验研究中,丙烯酸块中总共40个植入物(6?×10?×20≤mm)垂直于表面安装。然后将它们随机分为四组:(1)用30?ncm,(2)扭曲一次,用30℃和5分钟的间隔扭曲三次,(3)用30?ncm扭转一次,打开螺钉,打开螺丝用30?NCM重新缠绕,(4)用35厘米扭转一次。通过使用数字扭矩计确认扭矩值。然后,在测量水池测量值之前,样品在测量前三个小时接受了力(2?CPS,0.453-11.793Ω·kg)。然后测量并记录螺杆松动力(扭矩)。使用单向ANOVA和Tukey Hoc测试以5%的错误级别进行SPSS(版本22)软件分析所获得的数据。结果 。据报道,第4组(27.8?±1.3),1(26.8?±1.3),1(25.1?±1.3),最大平均螺杆的最大平均水坝值在第2组(24.9?±1.2)中发现了平均排水量值。此外,在第2和3()之间没有观察到显着差异,但在第1和3组和其他组()之间发现了显着差异。结论 。扭矩值的增加增加了扭矩损耗。然而,第4组的Detorque值显示出与制造商推荐的价值的最小差异(30?NCM)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号