System Accuracy Assessment of a Combined Invasive and Noninvasive Glucometer

机译:组合侵入性和非侵入性血糖仪的系统精度评估

获取原文
           

摘要

Background: The pain associated with pricking the fingertip for blood glucose self-testing is considered to be a major burden in diabetes treatment. This study was performed to evaluate the system accuracy of the invasive TensorTip Combo Glucometer (CoG) device component in accordance with ISO15197:2015 requirements and to explore the accuracy of the noninvasive tissue glucose prediction component. Methods: One hundred samples were obtained from people with type 1 and type 2 diabetes and healthy volunteers (43 females, 57 males; age: 53 ± 16 years), with glucose distribution as requested by the ISO standard. Three strip lots were tested twice by healthcare professionals in comparison to YSI 2300 Stat Plus reference method followed by a noninvasive tissue glucose reading (NI-CoG). Mean Absolute (Relative) Difference (MARD) was calculated and a consensus error grid (CEG) analysis was performed. Results: The ISO system accuracy criteria were met with the invasive strip technology by 586/600 of the data points (97.1%) and for each strip lot separately. All invasive results (100%) were within CEG-zone A and total MARD was calculated to be 7.1%. With the noninvasive reading, 99% of raw data points were in A + B (91.1% and 7.8%), and the total MARD was calculated to be 18.1%. Discussion: The invasive component of the CoG device was shown to be in full compliance with the current ISO15197 criteria. Good results were also obtained with the NI-CoG tissue glucose prediction. This noninvasive technology would potentially be suitable for frequent pain-free glucose monitoring in many people with diabetes.
机译:背景:与刺穿血糖自检相关的疼痛被认为是糖尿病治疗的主要负担。该研究进行了根据ISO15197:2015要求评估侵入式Tensortip Combo Glucometer(COG)装置组件的系统精度,并探讨非侵入式组织葡萄糖预测部件的准确性。方法:从1型和2型糖尿病和健康志愿者(43名女性,57名男性;年龄:53±16岁)的人中获得一百个样品,如ISO标准所要求的葡萄糖分布。与YSI 2300 STAT PLUS参考方法相比,医疗保健专业人员进行了两次通过医疗保健专业人员测试了三条批量批次。随后是非侵入式组织葡萄糖读数(Ni-Cog)。计算平均绝对(相对)差(MARD),并进行共识误差网格(CEG)分析。结果:ISO系统精度标准通过586/600的数据点(97.1%)和单独的每条剥离批次达到侵入性带技术。所有侵入性结果(100%)都在CEG区A中,并计算总狂欢为7.1%。随着非侵入性阅读,99%的原料数据点处于A + B(91.1%和7.8%),总稿是计算为18.1%。讨论:COG器件的侵入成分显示出完全符合当前ISO15197标准。 Ni-Cog组织葡萄糖预测也获得了良好的结果。这种非侵入性技术可能适用于许多糖尿病的频繁无痛苦的葡萄糖监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号