Background. This study was conducted to compare the cyclic fatigue resistance of VDW.ROTATE, TruNatomy Prime, HyFlex CM, and 2Shape nickel-titanium (NiTi) rotary instruments in double-curved canals in a simulated clinical environment. Methods. Eighty NiTi files were used for the fatigue testing in stainless steel canals compatible with instrument sizes until fracture occurred (n=20): VDW.ROTATE (#25/.04), TruNatomy Prime (#26/.04), HyFlex CM (#25/.04) and 2Shape TS1 (#25/.04). For each instrument, the number of cycles to fracture (NCF) was calculated, and the fractured fragment length (FL) was measured. To verify that the files were fractured due to cyclic fatigue, the fractured surfaces of the files were evaluated under a scanning electron microscope. Data were statistically analyzed using the Kruskal–Wallis and Student’s t-tests at the 95% confidence level. Results. The failure of the files due to cyclic fatigue was first seen in the apical curvature before the coronal curvature (P0.05). The highest fatigue resistance was observed in VDW.ROTATE and HyFlex CM files in both curvatures (P0.05). There were no significant differences in the fatigue resistance between the HyFlex CM and VDW.ROTATE files or between the 2Shape and the TruNatomy files (P0.05). There was no difference in the fractured lengths of the files between the apical and coronal curvatures (P0.05). Conclusion. In artificial S-shaped root canals, VDW.ROTATE and HyFlex CM files exhibited higher fatigue resistance compared to 2Shape and TruNatomy files.
机译:背景。进行了该研究以比较VDW.Rotate,Trunatomy Prime,Hyflex Cm和2形镍 - 钛(NITI)旋转仪器在模拟临床环境中的循环疲劳性。方法。八十个NITI档案用于与仪器尺寸兼容的不锈钢运河中的疲劳测试,直到发生骨折(n = 20):VDW.ROTATE(#25 / .04),Trunatomy Prime(#26 / .04),Hyflex cm( #25 / .04)和2shape ts1(#25 / .04)。对于每种仪器,计算裂缝(NCF)的循环次数,并测量裂缝片段长度(FL)。为了验证由于循环疲劳而裂缝的文件,在扫描电子显微镜下评估文件的裂缝表面。使用Kruskal-Wallis和学生的T-Tests在95%的置信水平上进行统计分析数据。结果。在冠状曲率之前首先在顶端曲率中看到由于循环疲劳引起的文件的失败(P <0.05)。在VDW.ROTATE和HYFLEX CM文件中观察到最高的疲劳阻力在两个曲率下(P <0.05)。 HYFlex CM和VDW.ROTATE文件之间或2SHAPE与TRURUTOMY文件之间没有显着差异,或者在2SHAPE和TRURURATOMY文件之间(P> 0.05)之间。在顶端和冠状曲率之间的文件的骨折长度没有差异(P> 0.05)。结论。在人造的根部管中,VDW.ROTATE和HYFLEX CM文件与2SSHAPE和TRURUTOMY文件相比表现出更高的疲劳电阻。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号