A Contact-Free, Ballistocardiography-Based Monitoring System (Emfit QS) for Measuring Nocturnal Heart Rate and Heart Rate Variability: Validation Study

机译:基于接触的颅骨造影的监测系统(EMFIT QS),用于测量夜间心率和心率变异性:验证研究

获取原文
           

摘要

Background: Heart rate (HR) and heart rate variability (HRV) measurements are widely used to monitor stress and recovery status in sedentary people and athletes. However, effective HRV monitoring should occur on a daily basis because sparse measurements do not allow for a complete view of the stress-recovery balance. Morning electrocardiography (ECG) measurements with HR straps are time-consuming and arduous to perform every day, and thus compliance with regular measurements is poor. Contact-free, ballistocardiography (BCG)-based Emfit QS is effortless for daily monitoring. However, to the best of our knowledge, there is no study on the accuracy of nocturnal HR and HRV measured via BCG under real-life conditions. Objective: The aim of this study was to evaluate the accuracy of Emfit QS in measuring nocturnal HR and HRV. Methods: Healthy participants (n=20) completed nocturnal HR and HRV recordings at home using Emfit QS and an ECG-based reference device (Firstbeat BG2) during sleep. Emfit QS measures BCG by a ferroelectret sensor installed under a bed mattress. HR and the root mean square of successive differences between RR intervals (RMSSD) were determined for 3-minute epochs and the sleep period mean. Results: A trivial mean bias was observed in the mean HR (mean –0.8 bpm [beats per minute], SD 2.3 bpm, P =.15) and Ln (natural logarithm) RMSSD (mean –0.05 ms, SD 0.25 ms, P =.33) between Emfit QS and ECG. In addition, very large correlations were found in the mean values of HR ( r =0.90, P .001) and Ln RMSSD ( r =0.89, P .001) between the devices. A greater amount of erroneous or missing data ( P .001) was observed in the Emfit QS measurements (28.3%, SD 14.4%) compared with the reference device (1.1%, SD 2.3%). The results showed that 5.0% of the mean HR and Ln RMSSD values were outside the limits of agreement. Conclusions: Based on the present results, Emfit QS provides nocturnal HR and HRV data with an acceptable, small mean bias when calculating the mean of the sleep period. Thus, Emfit QS has the potential to be used for the long-term monitoring of nocturnal HR and HRV. However, further research is needed to assess reliability in HR and HRV detection.
机译:背景:心率(HR)和心率变异性(HRV)测量被广泛用于监测久坐人和运动员中的压力和恢复状态。但是,有效的HRV监测应每天进行,因为稀疏测量不允许完全查看压力恢复平衡。早晨心电图(ECG)测量与HR带子每天都耗时,表现艰巨,因此遵守定期测量差。无与伦比的芭蕾舞通知(BCG)的EMFIT QS是日常监测的毫不费力。然而,据我们所知,没有关于通过BCG在现实条件下通过BCG测量的夜间HR和HRV的准确性的研究。目的:本研究的目的是评估EMFIT QS测量夜间HR和HRV的准确性。方法:健康参与者(n = 20)在睡眠期间使用EMFIT QS和基于ECG的参考装置(FirstBeat BG2)在家完成夜间人力资源和HRV录像。 EMFIT QS通过安装在床垫下的erdroelectret传感器测量BCG。 HR和RR间隔(RMSD)之间连续差异的均方根均为3分钟的时期,睡眠周期意味着。结果:在平均值的平均值(平均值-0.8bpm =每分钟)中观察到琐碎的平均偏见,SD 2.3 BPM,P = .15)和LN(自然对数)RMSD(平均值-0.05ms,SD 0.25毫秒,P EMFIT QS和ECG之间的= .33)。此外,在器件之间的HR(r = 0.90,p <.001)和LN RMSD(r = 0.89,p <.001)的平均值中发现了非常大的相关性。与参考装置(1.1%,SD 2.3%)相比,在EMFIT QS测量中观察到更大的错误或缺失的数据(P <.001)(28.3%,SD 14.4%)。结果表明,5.0%的平均HR和LN RMSD值超出了协议的范围。结论:根据目前的结果,EMFIT QS在计算睡眠周期的平均值时提供可接受的小平均偏差的夜间HR和HRV数据。因此,EMFIT QS有可能用于夜间HR和HRV的长期监测。然而,需要进一步研究以评估HR和HRV检测的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号