首页> 外文期刊>Endocrine journal >Factory-calibrated continuous glucose monitoring and capillary blood glucose monitoring in a case with insulinoma: usefulness and possible pitfall under chronic hyperinsulinemic hypoglycemia
【24h】

Factory-calibrated continuous glucose monitoring and capillary blood glucose monitoring in a case with insulinoma: usefulness and possible pitfall under chronic hyperinsulinemic hypoglycemia

机译:在胰岛素瘤的情况下,工厂校准连续葡萄糖监测和毛细血糖监测:慢性高胰岛素血症低血糖症下的有用性和可能的​​缺陷

获取原文
       

摘要

The accuracy of factory-calibrated continuous glucose monitoring (fCGM) within hypoglycemic ranges, especially under the status of chronic hyperinsulinemic hypoglycemia like insulinomas, remains an issue. Even so, fCGM is known to be useful for detecting hypoglycemia unawareness in insulinoma cases. A 25-year-old woman presenting with sudden unconsciousness was diagnosed with insulinoma; fCGM facilitated diagnosis by continuous monitoring for hypoglycemia. Before surgery, she was treated with continuous and frequent bolus infusions of 50% glucose via central venous catheter. To evaluate the accuracy of fCGM values in this case, a comparison between fCGM and capillary blood glucose (CBG) values was also performed. According to the simultaneously measured values, those of fCGM were largely in accordance with those of CBG. Moreover, compared with the previously reported case not having glucose infusions via central venous catheter, both the mean absolute relative differences (MARDs) and the absolute differences (Δ glucose) between fCGM and CBG values were larger in the present case, although no significant differences of MARDs and Δ glucose between the two cases were observed in several different conditions including fasting, post-meal, hypoglycemia, and others. Therefore, we should note possible increased differences between fCGM and CBG values in cases using frequent intravenous glucose infusions as well as case-dependent differing levels of consistency between them.
机译:降血糖范围内的工厂校准连续葡萄糖监测(FCGM)的准确性,特别是在胰岛素瘤等慢性高胰岛素血症低血基血症的状态下仍然存在问题。即便如此,已知FCGM可用于检测胰岛素壳体中的低血糖无意。一名25岁的女性患有突然的无意识患有胰岛素瘤; FCGM通过连续监测进行低血糖症促进诊断。在手术前,她通过中央静脉导管用连续和频繁的葡萄糖输注进行治疗。为了评估FCGM值的准确性在这种情况下,还进行了FCGM和毛细血管血糖(CBG)值之间的比较。根据同时测量值,FCGM的那些主要根据CBG的价值。此外,与先前报道的病例相比,没有通过中央静脉导管的葡萄糖输注,在当前情况下,FCGM和CBG值之间的平均绝对相对差异(MARCE)和绝对差异(Δ葡萄糖)虽然没有显着差异在几种情况下,在几种不同的条件下观察到两种情况之间的狂欢和血糖,包括禁食,膳食膳食,低血糖和其他情况。因此,我们应该注意在使用频繁的静脉内葡萄糖输注的情况下,FCGM和CBG值之间可能增加的差异以及它们之间的案例相关的不同程度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号