>Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum
【24h】
Application of Molecularly Imprinted Polymer-modified Potentiometric Sensor for Quantitative Determination of Histamine in Serum
A molecularly imprinted polymer-modified potentiometric histamine (HIS) sensor was prepared and used for quantitative determination of HIS in bovine serum. The calibration curve using the potential responses measured in 1 × 10~(?3) mol L~(?1) phosphate buffer (pH 7.4) showed good linearity in the HIS concentration range of 3 × 10~(?4) to 1 × 10~(?2) mol L~(?1) (r = 0.92), with a detection limit of 1.6 × 10~(?4) mol L~(?1). In bovine serum samples, the HIS sensor showed good recovery values of 91 – 104%. Therefore, this HIS sensor successfully determined the HIS concentration in bovine serum samples.
展开▼
机译:制备分子印迹聚合物改性的聚合物组胺(HIS)传感器并用于定量测定他牛血清的测定。使用在1×10〜(α3)摩尔L〜(α1)磷酸盐缓冲液(pH7.4)中测量的潜在响应的校准曲线在其浓度范围内为3×10〜(Δ4)至1×,显示出良好的线性。 10〜(?2)mol L〜(α1)( r = 0.92),检测限为1.6×10〜(Δ4)mol l〜(α1)。在牛血清样品中,他的传感器显示出91-104%的良好恢复值。因此,他的传感器成功地确定了他在牛血清样品中的浓度。