To experimentally compare the optical performance of two different Extended-Depth-of-Focus (EDOF) intraocular lenses (IOLs) using a standardized optical bench set-up. In this experimental study, following IOLs were assessed: the TECNIS? Symfony ZXR00 (Johnson&Johnson, Santa Ana, USA) and the AT LARA 829MP (Carl Zeiss Meditec, Jena, Germany) IOLs. The through-focus modulation transfer function (MTF) values were measured at a spatial frequency of 50 lp/mm and at aperture sizes of 2, 3, and 4.5?mm. Each IOL was measured while centered using ISO 11979-2 Model 1 (aberration-free) and Model 2 (+?0.28?μm spherical aberration) corneas. United States Air Force (USAF) target images were also recorded for a qualitative evaluation. At 2?mm pupil with ISO1 cornea, the primary and secondary foci of both IOLs appeared to merge, providing an elongated depth of focus. At 3 and 4.5?mm pupil sizes, the through-focus MTF curves of both IOLs showed a bifocal-like V-pattern. While the Symfony IOL showed an overall superior MTF values when measured with the ISO2 cornea, the opposite propensity could be observed with the AT LARA IOL. This optical behavior could be qualitatively confirmed by the USAF target images. Although the two EDOF IOLs share similarities in their optical properties, the main difference lies in their optical design and performance with respect to spherical aberration. Such characteristics should be taken into account during IOL and patient selection.
机译:通过标准化的光学台面设定实验地比较两种不同延伸深度(EDOF)眼镜透镜(IOL)的光学性能。在这个实验研究中,在评估IOL之后:TECNIS? Symfony ZXR00(Johnson&Johnson,Santa Ana,USA)和在Lara 829MP(Carl Zeiss Meditec,Jena,德国)IOL。通过50磅/mm的空间频率和2,3和4.5Ωmm的空间频率测量通过聚焦调制传递函数(MTF)值。使用ISO 11979-2型号1(无像差)和模型2(+0.28≤μm球面像差)射击每个IOL。美国空军(USAF)目标图像也被记录进行定性评估。用ISO1角膜2瞳孔,两个IOL的初级和二级灶具似乎合并,提供了细长的焦深。在3和4.5?MM瞳孔尺寸下,两个IOL的通焦MTF曲线显示了双焦点的V形型。虽然Symfony IOL在用ISO2角膜测量时显示了整体优越的MTF值,但可以在Lara IOL中观察到相反的倾向。 USAF目标图像可以定性地确认该光学行为。虽然两个EDOF IOL在光学性质中共享相似性,但主要差异在于它们的光学设计和性能相对于球面像差。在IOL和患者选择期间,应考虑这些特性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号