Introduction: Capnometry is not well studied in neonates on Patient Triggered Ventilation (PTV). We conducted this study to determine the reliability of Main Stream (MS) and Side Stream (SS) Capnometry in neonates on PTV.Method: Neonates on PTV were enrolled in the study. Before each Arterial Blood Gas (ABG) assessment, MS and SS capnometer readings were recorded. Lung mechanics were recorded from the online graphics. ABG was collected from the arterial line. The EtCO2 ? PaCO2 correlation was drawn for MS and SS capnometry under various disease conditions, ventilator settings and lung mechanics.Results: A total of 74 ABGs were collected from 18 patients. The EtCO2 ? PaCO2 correlation was better with MS capnometry than with SS capnometry (r = 0.855 vs.0.68, p 60 /min), SS capnometry correlated poorly. The EtCO2 ?PaCO2 correlation by both methods was better when lung compliance was > 1 ml / cmH2O than when lung compliance was < 1 ml / cmH2O.Conclusions: MS capnometry is superior to SS capnometry for neonates on patient triggered ventilation.
机译:简介:对于患者触发式通气(PTV),新生儿的二氧化碳监测没有得到很好的研究。我们通过这项研究来确定PTV新生儿的主流(MS)和旁流(SS)二氧化碳测定法的可靠性。方法:将PTV上的新生儿纳入研究。在每次进行动脉血气(ABG)评估之前,都要记录MS和SS二氧化碳检测仪读数。肺力学记录来自在线图形。 ABG是从动脉线收集的。 EtCO2?在各种疾病状况,呼吸机设置和肺力学情况下,MS和SS二氧化碳描记法均绘制了PaCO2相关性。结果:从18例患者中共收集到74种ABG。 EtCO2? MS二氧化碳测定法的PaCO2相关性优于SS二氧化碳测定法(r = 0.855 vs.0.68,p 60 / min),SS二氧化碳测定法相关性较差。当肺部顺应性> 1 ml / cmH2O时,两种方法的EtCO2?PaCO2相关性要好于肺部顺应性<1 ml / cmH2O时。结论:对于患者触发通气的新生儿,MS二氧化碳测定法优于SS二氧化碳测定法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号