首页> 外文期刊>Journal of Clinical Microbiology >Interpretive criteria and quality control limits for ceftibuten disk susceptibility tests. Collaborative Antimicrobial Susceptibility Testing Group.
【24h】

Interpretive criteria and quality control limits for ceftibuten disk susceptibility tests. Collaborative Antimicrobial Susceptibility Testing Group.

机译:头孢替丁片药敏试验的解释性标准和质量控制限制。协同药敏试验组。

获取原文
           

摘要

In vitro studies were undertaken to evaluate susceptibility tests with 30-micrograms ceftibuten disks. The following interpretive criteria were proposed: less than or equal to 17 mm for resistance (MIC, greater than or equal to 32 micrograms/ml) and greater than or equal to 21 mm for susceptibility (MIC, less than or equal to 8.0 micrograms/ml). A multilaboratory quality control study led to the conclusion that Escherichia coli ATCC 25922 should provide zones 29 to 35 mm in diameter.
机译:进行了体外研究,以评估使用30毫克头孢替丁圆片的药敏试验。提出以下解释性标准:电阻小于或等于17毫米(MIC,大于或等于32微克/毫升)和磁化率大于或等于21毫米(MIC,小于或等于8.0微克/毫升)。毫升)。一项多实验室质量控制研究得出的结论是,大肠杆菌ATCC 25922应当提供直径29至35毫米的区域。