首页> 外文期刊>Journal of Clinical Microbiology >Comparison of radioimmunoassay and fluorescent polarization immunoassay for quantitative determination of vancomycin concentrations in serum.
【24h】

Comparison of radioimmunoassay and fluorescent polarization immunoassay for quantitative determination of vancomycin concentrations in serum.

机译:放射免疫分析法与荧光偏振免疫分析法用于定量测定血清中万古霉素浓度的比较。

获取原文
           

摘要

A new fluorescent polarization immunoassay (Abbott Laboratories, Diagnostics Division, North Chicago, Ill.) was compared with a standard radioimmunoassay (American Diagnostics Corp., Newport Beach, Calif.) in 34 patients being treated with vancomycin. A total of 123 serum samples were divided and quantitatively analyzed for vancomycin by both assay methods. The results obtained indicated that the two assay methods are comparable (y = 1.01x - 0.81; r = 0.99).
机译:在接受万古霉素治疗的34例患者中,将一种新的荧光偏振免疫分析法(伊利诺伊州北芝加哥诊断分部的Abbott实验室)与标准放射免疫分析法(加利福尼亚州纽波特比奇的美国诊断公司)进行了比较。总共分出123个血清样品,并通过两种测定方法对万古霉素进行了定量分析。获得的结果表明两种测定方法是可比较的(y = 1.01x-0.81; r ​​= 0.99)。