Comparison of Three Intraocular Pressure Measurement Methods Including Biomechanical Properties of the Cornea

机译:包括角膜生物力学特性在内的三种眼压测量方法的比较

获取原文
           

摘要

Purpose.: The aim of this study was to show the usefulness of three methods for measuring IOP: Goldmann applanation tonometry, rebound tonometry, and Ultra-High-Speed Scheimpflug technology. Methods.: The examined group consisted of 96 patients (192 eyes), including 63 women and 33 men with a mean age of 59.3 ?± 19.9 years. Together, 152 healthy eyes and 40 eyes with different pathologies were examined. Intraocular pressure was measured using the Goldmann applanation tonometer (GAT), the Icare Pro rebound tonometer (RT), and Ultra-High-Speed Scheimpflug technology (UHS ST; Corvis ST with pachymetry). Additionally, corneal pachymetry was conducted with a Scheimpflug camera (Pentacam) and an Ultrasound Pachymeter (A-scan Plus) as a comparison for Corvis ST pachymetry. Results.: The mean IOPs were 15.6 ?± 3.75 mm Hg, 15.6 ?± 3.5 mm Hg, and 16.1 ?± 4.0 mm Hg when measured with the GAT, the RT, and the UHS ST, respectively. The mean central corneal thickness (CCT) was 543.7 ?± 52.7 ??m, 547.9 ?± 54.0 ??m, and 556.25 ?± 38.8 ??m as measured with the UHS ST, the Pentacam, and the Ultrasound Pachymeter, respectively. In comparison between devices, there was a significant difference between IOP values measured with the GAT and the RT versus the UHS ST (P 0.001), and there was no significant difference between GAT and RT (P = 0.5). No significant differences were observed in CCT measured with the UHS ST, Pentacam, and Ultrasound Pachymeter. Conclusions.: We showed that the RT Icare Pro ensures IOP measurements that are more comparable with the measurements obtained with the GAT than the measurements that are provided by UHS ST.
机译:目的:本研究的目的是展示三种测量IOP的方法的实用性:Goldmann压平眼压测量,回弹眼压测量和超高速Scheimpflug技术。方法:检查组由96例患者(192眼)组成,包括63名女性和33名男性,平均年龄为59.3±19.9岁。总共检查了152眼健康的眼睛和40眼具有不同病理的眼睛。使用Goldmann压平眼压计(GAT),Icare Pro回弹眼压计(RT)和超高速Scheimpflug技术(UHS ST;带测厚仪的Corvis ST)测量眼压。另外,用Scheimpflug相机(Pentacam)和超声测厚仪(A-scan Plus)进行角膜测厚,作为Corvis ST测厚的比较。结果:当用GAT,RT和UHS ST测量时,平均IOP分别为15.6±3.75 mm Hg,15.6±±3.5 mm Hg和16.1±±4.0 mm Hg。用UHS ST,Pentacam和超声测厚仪测得的平均中央角膜厚度(CCT)分别为543.7±52.7μm,547.9±54.0μm和556.25±38.8μm。在设备之间进行比较时,用GAT和RT测得的IOP值与UHS ST之间存在显着差异(P <0.001),而GAT和RT之间没有显着差异(P = 0.5)。用UHS ST,Pentacam和超声测厚仪测得的CCT没有观察到显着差异。结论:我们证明,RT Icare Pro可确保IOP测量值与使用GAT获得的测量值相比,可比UHS ST提供的测量值更具可比性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号