首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Effects of Hemolysis on the Roche Ammonia Method for Hitachi Analyzers
【24h】

Effects of Hemolysis on the Roche Ammonia Method for Hitachi Analyzers

机译:溶血对日立分析仪罗氏氨法的影响

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The manufacturer’s product literature for the Roche ammonia reagent (cat. no. 11877984) states that “there is no interference from hemolysis up to a hemolysis index of 50 (approximate hemoglobin concentration: 50 mg/dL)”. This statement can and has been interpreted to mean there are artifactual changes in the ammonia results above this hemolysis index limit. The likely result is that laboratories could use this low limit as the definitive guide and not report any results above it. This would lead to rejection of up to 10% of samples.The effect of hemolysis on analytes is often studied [e.g., Ref. (1 …
机译:制造商的罗氏氨试剂的产品文献(目录号11877984)指出:“溶血指数达到50(血红蛋白浓度约为50 mg / dL)时,不会发生溶血干扰”。该陈述可以并且已经被解释为意味着高于该溶血指数限值的氨结果存在人为变化。可能的结果是,实验室可以将这个下限用作确定的指南,而不会报告超出该上限的任何结果。这将导致最多10%的样品被拒。经常研究溶血对分析物的影响[例如,参考文献(1…