摘要

Background: We explored the potential of a microfluidic device based on centrifugal force as an immunoassay platform by examining the imprecision of assays carried out with 200 nL of sample.Methods: Biotinylated antibodies against α-fetoprotein (AFP), interleukin-6 (IL-6), and carcinoembryonic antigen [(CEA); 0.1 g/L each in 15 mmol/L phosphate-buffered saline (PBS) containing 0.1 mL/L Tween 20] were attached to a microcolumn packed with streptavidin-coated particles. A 200-nL sample was then allowed to pass through the microcolumn for 240 s, followed by Alexa 647–labeled detection antibody (7.5 mg/L in 15 mmol/L PBS containing 10 g/L bovine serum albumin). The flow rate was controlled by altering the rotational speed. Up to 104 sandwich type immunoassays were completed within 50 min.Results: For AFP, IL-6, and CEA the detection limits were, respectively, 0.15, 1.25, and 1.31 pmol/L. Inter- and intraassay imprecisions (CVs) were 10% and 20%, respectively, for analyte concentrations 5 pmol/L. The CEA antibody had the lowest affinity according to fluorescence image analysis of the microcolumn region. The result was confirmed in a comparative study using BIAcore 3000.Conclusions: Day-to-day (total) imprecision (CV) of immunoassays on the compact disc–shaped device are 20%. Analysis of fluorescence images allows rapid ranking of antibodies according to their affinities.
机译:背景:我们通过检查200 nL样品的不精确性,探索了基于离心力的微流体装置作为免疫测定平台的潜力。方法:针对甲胎蛋白(AFP),白介素6(IL- 6)和癌胚抗原[(CEA);将含有0.1 mL / L吐温20]的15 mmol / L磷酸盐缓冲液(PBS)中的0.1 g / L分别连接到装有链霉亲和素包被颗粒的微柱上。然后使200 nL样品通过微柱240 s,然后通过Alexa 647标记的检测抗体(7.5 mg / L的15 mmol / L PBS中含有10 g / L牛血清白蛋白)。通过改变转速来控制流量。在50分钟内最多完成104次夹心型免疫分析。结果:对于AFP,IL-6和CEA,检测限分别为0.15、1.25和1.31 pmol / L。对于分析物浓度> 5 pmol / L,测定间和测定内不精密度(CV)分别为<10%和<20%。根据微柱区域的荧光图像分析,CEA抗体的亲和力最低。使用BIAcore 3000进行的比较研究证实了该结果。结论:光盘状设备上的免疫测定的每日(总)不精确度(CV)小于20%。荧光图像分析允许根据抗体的亲和力对抗体进行快速排名。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号