首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Defects in diagnostic kits for determination of urate in serum.
【24h】

Defects in diagnostic kits for determination of urate in serum.

机译:诊断试剂盒中用于确定血清中尿酸盐的缺陷。

获取原文
           

摘要

Certain diagnostic kits that measure serum urate by the Barham and Trinder principle of enzymic liberation of oxygen and its combination with chromogens can give results for urate in fresh serum that are approximately 20% lower than results from serum stored at ambient temperature for 72 h. In fresh serum, antioxidants compete with chromogen for liberated peroxyl-oxygen. We postulate that during storage the interfering antioxidant substances are destroyed. In some diagnostic kits, L-ascorbate oxidase is added to the reaction, eliminating some but not all of this effect. We discuss defects of several commercially available kits for determination of serum urate and recommend comparing results of these kits with results from the phosphotungstic acid method as a precaution against falsely low results.
机译:某些通过Barham和Trinder的氧气酶解原理以及其与色原的组合来测量血清尿酸盐的诊断试剂盒,可以得出新鲜血清中尿酸盐的结果,其结果要比在室温下储存72小时的血清结果低约20%。在新鲜的血清中,抗氧化剂与色原竞争释放出的过氧-氧。我们假设在存储过程中会破坏干扰性​​抗氧化剂物质。在某些诊断试剂盒中,L-抗坏血酸氧化酶被添加到反应中,消除了部分但不是全部这种作用。我们讨论了几种用于测定血清尿酸盐的市售试剂盒的缺陷,并建议将这些试剂盒的结果与磷钨酸方法的结果进行比较,以防出现假低结果。