Four fluorescence polarization immunoassays for therapeutic drug monitoring evaluated.

机译:评价了用于治疗药物监测的四种荧光偏振免疫测定法。

获取原文
           

摘要

We evaluated four fluorescence polarization immunoassays--those for phenytoin, procainamide, N-acetylprocainamide (NAPA), and quinidine--from Roche Diagnostic Systems, done in the Cobas Bio FP centrifugal analyzer. The assays for phenytoin, NAPA, and quinidine demonstrated a linear response over the expected ranges of concentrations, and analytical recovery of test drug added to drug-free sera was greater than 90%. However, recovery in the procainamide assay was poor (69-82%) for samples containing greater than 8 mg/L, owing to nonlinearity. Results of method-comparison studies of the four assays paralleled the recovery studies, although the quinidine assay demonstrated a bias (1.0 mg/L higher) when compared with EMIT (Syva). The precision of the phenytoin assay was acceptable at all concentrations tested (total CV less than 7.0%). Imprecision of the other assays was significant at certain concentrations: total CV greater than 10.0% at subtherapeutic values for NAPA and quinidine, and greater than 9.0% for low (2.4 mg/L) and moderate concentrations (9.6 mg/L) of procainamide. Interferences were not significant for hemolyzed, icteric, or lipemic specimens or for specimens with added drug metabolites. The calibration curves for all four assays had good stability (greater than 60 days).
机译:我们评估了Roche Diagnostic Systems的四种荧光偏振免疫测定法-苯妥英钠,普鲁卡因酰胺,N-乙酰procainamide(NAPA)和奎尼丁-在Cobas Bio FP离心分析仪中进行。苯妥英钠,NAPA和奎尼丁的测定结果表明,在预期的浓度范围内线性响应,添加到无药血清中的测试药物的分析回收率大于90%。但是,由于非线性,对于大于8 mg / L的样品,普鲁卡因酰胺测定的回收率很低(69-82%)。四种测定的方法比较研究结果与回收率研究相似,尽管与EMIT(Syva)相比,奎尼丁测定显示出偏倚(高1.0 mg / L)。在所有测试浓度下(总CV小于7.0%),苯妥英钠测定的精密度都是可以接受的。在某些浓度下,其他测定的不精确性很显着:NAPA和奎尼丁在亚治疗值下的总CV大于10.0%,而普鲁卡因酰胺的低浓度(2.4 mg / L)和中度浓度(9.6 mg / L)的总CV大于9.0%。对于溶血,黄疸或脂血症的标本或添加了药物代谢物的标本,干扰不明显。所有四种测定的校准曲线均具有良好的稳定性(大于60天)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号