首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Amino acid determination in biological fluids by automated ion-exchange chromatography: performance of Hitachi L-8500A
【24h】

Amino acid determination in biological fluids by automated ion-exchange chromatography: performance of Hitachi L-8500A

机译:自动离子交换色谱法测定生物体液中的氨基酸:Hitachi L-8500A的性能

获取原文
获取外文期刊封面目录资料

摘要

The Hitachi L-8500A is a newly available apparatus for amino acid (AA) analysis that allows automatic on-line mixing of the ninhydrin reagent. The within-run precision (human plasma pools at three different concentrations) showed CVs 3.8% except for the lowest concentration of citrulline (4.4%), Tyr (4.5%), and α-aminobutyric acid (7.6%), and for the intermediate concentration of Asp (8.7%). Between-run precision (CV) was 3.1% for 17 AAs and 8.0% for 24 of 25 AAs (CV Asp = 12.0%). For retention times, within-run precision was 0.4% and between-run precision 1.8%. Excellent relations were found between the results from the Hitachi L-8500A and the widely used Beckman 6300 analyzer (0.929≤ r ≤0.999). The detection was still linear at 5 μmol/L except for Pro and hydroxyproline (20 μmol/L). The upper limit was at least 2500 μmol/L for 13 AAs and at least 1000 μmol/L for 27 of 29 AAs (anserine = 500, Val = 600 μmol/L). Values from 100 human plasma samples agreed with previously published data. We conclude that the results obtained with the Hitachi L-8500A are satisfactory when compared with those of other AA analyzers utilizing the same method. Furthermore, the Hitachi L-8500A displays several advantages including programming flexibility, microsample capacity, low noise plotting, ammonia filtering, and manual repacking of the analytical column.
机译:Hitachi L-8500A是一种新近用于氨基酸(AA)分析的仪器,可自动在线混合茚三酮试剂。批内精密度(三种不同浓度的人血浆池)显示CVs <3.8%,但瓜氨酸(4.4%),Tyr(4.5%)和α-氨基丁酸(7.6%)的最低浓度除外。中等浓度的Asp(8.7%)。对于17个AA,运行间精度(CV)为<3.1%,对于25个AA中的24个,运行间精度(CV)为<8.0%(CV Asp = 12.0%)。对于保留时间,批内精密度<0.4%,批间精密度<1.8%。日立L-8500A的结果与广泛使用的Beckman 6300分析仪之间的关系极佳(0.929≤r≤0.999)。除Pro和羟脯氨酸(20μmol/ L)外,在5μmol/ L处检测仍呈线性。 13个AA的上限至少为2500μmol/ L,29个AA中的27个的上限为至少1000μmol/ L((色= 500,Val = 600μmol/ L)。来自100个人体血浆样品的值与以前发表的数据一致。我们得出的结论是,与使用相同方法的其他AA分析仪相比,Hitachi L-8500A获得的结果令人满意。此外,日立L-8500A还具有多种优势,包括编程灵活性,微量样品容量,低噪声绘图,氨气过滤和分析柱的手动重新装箱。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号