首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Graphite furnace atomic absorption spectroscopic measurement of blood lead in matrix-matched standards.
【24h】

Graphite furnace atomic absorption spectroscopic measurement of blood lead in matrix-matched standards.

机译:石墨炉原子吸收光谱法测定基体匹配标准品中的血铅。

获取原文
           

摘要

Now that the level of concern for a toxic blood lead concentration is 0.482 mumol/L (10 micrograms/dL), laboratories must meet new requirements to shorten analysis times and increase accuracy and precision of blood lead determinations. We used a matrix-matching method to estimate the lead concentration in blood by graphite furnace atomic absorption spectroscopy (GFAAS). For CDC proficiency samples and the NIST-Certified Blood Reference standard, the performance of this method compared favorably with that of previously published GFAAS methods and of the anodic stripping voltammetric method routinely used in our laboratory. At lead concentrations of 0.242 mumol/L (5.01 micrograms/dL) and 1.478 mumol/L (30.63 micrograms/dL), within-run CVs were 2.78% and 0.68%, respectively; between-run CVs were 4.9% and 1.35%. In 52 study samples with lead content ranging from 0.097 to 3.812 mumol/L (2 to 79 micrograms/dL), 87% of results by the matrix-modified method were within 0.048 mumol/L (1 microgram/dL) of consensus values.
机译:现在,有毒血铅浓度的关注水平为0.482摩尔/升(10微克/分升),实验室必须满足新的要求,以缩短分析时间并提高血铅测定的准确性和准确性。我们使用矩阵匹配方法通过石墨炉原子吸收光谱法(GFAAS)估算血液中的铅浓度。对于CDC能力样本和NIST认证的血液参考标准品,该方法的性能优于以前发布的GFAAS方法和我们实验室中常规使用的阳极溶出伏安法。铅浓度为0.242μmol/ L(5.01微克/ dL)和1.478μmol/ L(30.63微克/ dL)时,批内CV分别为2.78%和0.68%;运行间CV分别为4.9%和1.35%。在52个铅含量范围为0.097至3.812 mumol / L(2至79微克/ dL)的研究样本中,采用基质修饰方​​法的结果的87%在共识值的0.048 mumol / L(1微克/ dL)以内。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号