首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Heterologous double-determinant immunoradiometric assay CA 125 II: reliable second-generation immunoassay for determining CA 125 in serum.
【24h】

Heterologous double-determinant immunoradiometric assay CA 125 II: reliable second-generation immunoassay for determining CA 125 in serum.

机译:异源双重测定免疫放射测定法CA 125 II:可靠的第二代免疫测定法,用于测定血清中的CA 125。

获取原文
           

摘要

The new CA 125 II (Centocor) serum assay utilizing the M11 mouse monoclonal antibody as capture antibody and OC125 as tracer antibody, was investigated for its technical and clinical performance against the original CA 125 assay. The CA 125 II test revealed a quadruple increase in signal-to-noise ratio, good intra- and interassay precision (with CVs 5% and 7%, respectively), improved dilution linearity, and a minimal detectable dose of 0.38 units/mL. Sera were obtained from healthy females (n = 192), women with benign conditions (n = 208), and patients with various cancers (n = 379). Both assays measured highly similar CA 125 distributions with equal reference ranges and nearly identical positivity ( 35 units/mL) rates, resulting in similar receiver-operating characteristic curves and monitoring graphs. Linear regression analysis of results by the two assays (CA 125 = x, CA 125 II = y) in ovarian cancer patients showed, for CA 125 assay values between 0 and 1000 units/mL, a slope of 1.00 and a y-intercept of 12.6 (n = 254, r = 0.8617, P 0.0001). The heterologous CA 125 II assay appeared to be more accurate in patients who had human anti-mouse antibodies after immunoscintigraphy. The CA 125 II immunoradiometric assay is sensitive and reliable for measuring serum CA 125, and fully retains the cutoff values of 35 and 65 units/mL that were defined with the original CA 125 immunoradiometric assay.
机译:研究了使用M11小鼠单克隆抗体作为捕获抗体和OC125作为示踪抗体的新型CA 125 II(Centocor)血清测定相对于原始CA 125测定的技术和临床性能。 CA 125 II测试显示信噪比提高了四倍,批内和批间精密度均好(CV分别<5%和7%),稀释线性提高,最小可检测剂量为0.38单位/ mL 。血清来自健康的女性(n = 192),患有良性疾病的女性(n = 208)和患有各种癌症的患者(n = 379)。两种测定方法均以相同的参考范围和几乎相同的阳性率(> 35单位/ mL)速率测量了高度相似的CA 125分布,从而产生了相似的受体工作特征曲线和监测图。在卵巢癌患者中通过两种测定法(CA 125 = x,CA 125 II = y)对结果的线性回归分析表明,对于介于0和1000单位/ mL之间的CA 125测定值,斜率是1.00,y截距是12.6(n = 254,r = 0.8617,P <0.0001)。异种CA 125 II检测对于免疫闪烁扫描后具有人抗小鼠抗体的患者似乎更为准确。 CA 125 II免疫放射测定法对血清CA 125的测定是灵敏可靠的,并且完全保留了原始CA 125免疫放射测定法定义的35和65单位/ mL的临界值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号