Evaluation of accuracy of 20 different test kits for the enzymic determination of cholesterol.

机译:评估20种不同试剂盒的酶法测定胆固醇的准确性。

获取原文

摘要

We evaluated the accuracy of 20 different test kits for the enzymic determination of cholesterol. Using a selected set of standards, we found mean values differing by -9.9 to + 10.7% from values obtained by the Abell-Kendall procedure. Nine kits gave values with a relative bias of less than 2.5%. In the case of the other 11 kits, recalculation with secondary serum standards resulted in values within 5% of the reference values. Depending on the kit used, two different commercially available standard solutions produced results differing by, at most, 8.7%. We obtained indirect evidence that use of esterase of bacterial origin is associated with low values. We ascribe the high values obtained with some kits to positive interference by stabilizers or unspecified compounds in the reagents.
机译:我们评估了20种不同试剂盒的酶法测定胆固醇的准确性。使用一组选定的标准,我们发现平均值与通过Abell-Kendall程序获得的值相差-9.9至+ 10.7%。九种试剂盒给出的值的相对偏差小于2.5%。对于其他11种试剂盒,使用二级血清标准品重新计算得出的值在参考值的5%以内。根据所使用的套件,两种不同的市售标准溶液产生的结果相差最多为8.7%。我们获得了间接证据,证明细菌来源的酯酶的使用与低价值有关。我们将某些试剂盒获得的高值归因于试剂中稳定剂或未指定化合物的正干扰。