首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Lactate dehydrogenase isoenzyme 1 with the centrifugal analyzer after immunochemical removal of other isoenzymes.
【24h】

Lactate dehydrogenase isoenzyme 1 with the centrifugal analyzer after immunochemical removal of other isoenzymes.

机译:免疫化学去除其他同工酶后,用离心分析仪测定乳酸脱氢酶同工酶1。

获取原文
           

摘要

We describe a centrifugal analyzer method for measuring the LD-1 isoenzyme of lactate dehydrogenase (EC 1.1.1.27) in serum, after immunochemical precipitation of the other four isoenzymes. Enzymic activity was measured kinetically at 30 degrees C with the pyruvate-to-lactate assay. The method for LD-1 was linear to 1000 U/L. The precision (CV) of the assay was 1.0--2.2% within-run and 3.1--4.5% day-to-day. The reference interval was 26--73 U/L (n = 51), corresponding to 21--35% of total LD activity.
机译:我们描述了一种离心分析仪方法,用于测定其他四种同工酶的免疫化学沉淀后,血清中乳酸脱氢酶(EC 1.1.1.27)的LD-1同工酶。用丙酮酸至乳酸盐测定法在30℃下动力学测定酶活性。 LD-1的方法线性至1000 U / L。批内分析的精密度(CV)为1.0--2.2%,日常运行为3.1--4.5%。参考间隔为26--73 U / L(n = 51),相当于总LD活性的21--35%。