首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Continuous-flow enzymatic determination of total serum cholesterol and method standardization with CDC-calibrated pooled sera.
【24h】

Continuous-flow enzymatic determination of total serum cholesterol and method standardization with CDC-calibrated pooled sera.

机译:连续流酶法测定总血清胆固醇,并用CDC校准的合并血清进行方法标准化。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

We compared Boehringer Mannheim's enzymatic kit for the continuous-flow (AutoAnalyzer II) determination of serum cholesterol with Technicon's N-24a extraction method. Results for patients' samples analyzed by the enzymatic method were higher than those by the comparison method. To evaluate accuracy in the cholesterol determinations, we enrolled the enzymatic method into the Center for Disease Control's (CDC) Lipid Standardization program. We calibrated the method by use of a pooled sera for which cholesterol content was assigned by CDC after analysis by their reference Abell-Kendall procedure. We discuss the difficulties with available calibration material and limitations in the application of some commercial control materials to the enzymatic cholesterol method. The continuous-flow variables, Michaelis-Menton constants, percent ester-hydrolase activity, and other factors affecting the performance of the enzyme-linked cholesterol method are evaluated. We believe pooled sera with an assigned value for cholesterol content is the best calibrator material.
机译:我们将Boehringer Mannheim的酶试剂盒与Technicon的N-24a提取方法连续(AutoAnalyzer II)测定血清胆固醇进行了比较。酶法分析患者样品的结果高于比较法。为了评估胆固醇测定的准确性,我们将酶法纳入了疾病控制中心(CDC)脂质标准化计划中。我们通过使用合并血清进行校准,该合并血清由CDC根据参考Abell-Kendall程序分析后确定的胆固醇含量。我们讨论了可用的校准材料存在的困难以及某些商业控制材料在酶促胆固醇方法中的应用局限性。评估了连续流量变量,Michaelis-Menton常数,酯水解酶活性百分比以及其他影响酶联胆固醇法性能的因素。我们认为,为胆固醇含量指定值的合并血清是最好的校准材料。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号