Two conventional serum thyroxine assays, run in separate laboratories, one by competitive protein binding and one by radioimmunoassay, were used to evaluate the automated ARIA II (Becton Dickinson Immunodiagnostics) serum thyroxine assay. Competitive protein binding as compared to ARIA II with 111 clinical serum samples gave a slope of 1.04 and a correlation coefficient of 0.94. The radioimmunoassay comparison to ARIA II with 53 clinical serum samples gave a slope of 1.05 and a correlation coefficient of 0.92. The ARIA II inter-assay coefficient of variation for 10 replicates of low, medium, and high thyroxine serum samples was 6.2, 6.0, and 2.9%, respectively, with an inter-assay coefficient of variation among 15 different assays of 15.5, 10.1, and 7.9%. The automated ARIA II, with a 2.2-min cycle per sample, gives results that compare well with those by manual methodology.
机译:两种常规的血清甲状腺素测定分别在单独的实验室中进行,一种通过竞争性蛋白结合,一种通过放射免疫测定,用于评估自动ARIA II(Becton Dickinson免疫诊断)血清甲状腺素测定。与ARIA II相比,具有111种临床血清样品的竞争性蛋白质结合具有1.04的斜率和0.94的相关系数。与ARIA II进行的放射免疫分析与53个临床血清样品的比较得出斜率为1.05,相关系数为0.92。低,中和高甲状腺素血清样品10次重复的ARIA II测定间变异系数分别为6.2%,6.0和2.9%,在15种不同测定之间的测定间变异系数分别为15.5、10.1,和7.9%。自动的ARIA II,每个样品有2.2分钟的循环,其结果与手动方法可比。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号