摘要

We report here on novel biosensors for the highly sensitive and specific enzymatic analysis of glycated hemoglobin (HbA1c), which we measured primarily to identify three‐month average plasma glucose concentrations. Hydrogen peroxide (H2O2) was generated in this enzymatic measuring of glycated hemoglobin from oxidative cleavage of fructosyl valine by fructosyl amino acid oxidase, and we then used the generated hydrogen peroxide as a reducing agent of gold (III) ion to synthesize gold (0) as a nanoparticle. We were able to measure the gold nanoparticles obtained from this novel approach using colorimetry under a UV–VIS spectrophotometer. For this colorimetric detection, we analyzed 96‐well plates for glycated hemoglobin, and this system worked very well, with a detection limit of 0.142% and coefficient of variation below 10%. Finally, this method can measure glycated hemoglobin in patients with diabetes.
机译:我们在此报告了一种新型生物传感器,用于糖化血红蛋白(HbA1c)的高灵敏度和特异性酶促分析,我们主要进行测量以鉴定三个月的平均血浆葡萄糖浓度。通过果糖基氨基酸氧化酶对果糖基缬氨酸的氧化裂解,通过酶促测定糖基化血红蛋白生成了过氧化氢(H2O2),然后我们将生成的过氧化氢用作金(III)离子的还原剂来合成金(0)作为纳米粒子。我们能够在UV-VIS分光光度计下使用比色法测量从这种新方法获得的金纳米颗粒。对于这种比色检测,我们分析了96孔板中的糖化血红蛋白,该系统运行良好,检测限为0.142%,变异系数低于10%。最后,这种方法可以测量糖尿病患者的糖化血红蛋白。