首页> 外文期刊>Bulletin of the Korean Chemical Society >Application of a Membrane Desolvator to the Analysis of Organic Solvents in Inductively Coupled Plasma Atomic Emission Spectrometry
【24h】

Application of a Membrane Desolvator to the Analysis of Organic Solvents in Inductively Coupled Plasma Atomic Emission Spectrometry

机译:膜脱溶剂剂在电感耦合等离子体原子发射光谱法中有机溶剂分析中的应用

获取原文
           

摘要

A micro porous PTFE membrane desolvator (MMD) was built and evaluated for the on-line removal of organic solvents to facilitate the determination of trace metal contaminants in the solvents by ICP-AES. Three organic solvents, isopropyl alcohol (IPA), methanol, and dimethy sulfoxide (DMSO) were studied. The MMD reduced organic solvent concentration in the sample aerosol stream by 82% to 89%, as indicated by monitoring C(I) emission. Net signal intensity of Fe, Al, and Cu was increasing with higher organic solvent concentration, with the rate of increase being solvent dependent. The signal intensities for Mg and Pb followed the trend with the C(I) signal. Changing the sweep gas flow rate affected the optimum signal intensity. Wine samples were analyzed by the method of standard addition. The concentrations of B, Al, and Mg were determined with a relative precision of less than 2.3%.
机译:建立了微孔PTFE膜脱溶剂剂(MMD),并对其有机溶剂的在线去除进行了评估,以利于用ICP-AES测定溶剂中的痕量金属污染物。研究了三种有机溶剂,异丙醇(IPA),甲醇和二甲基亚砜(DMSO)。 MMD将样品气溶胶流中的有机溶剂浓度降低了82%至89%,如监测C(I)排放所表明的。 Fe,Al和Cu的净信号强度随着有机溶剂浓度的增加而增加,其增加速率取决于溶剂。 Mg和Pb的信号强度随C(I)信号的变化趋势而变化。改变吹扫气流速会影响最佳信号强度。通过标准添加方法分析葡萄酒样品。测定的B,Al和Mg的浓度相对精度小于2.3%。