首页> 外文期刊>BioMed research international >Left Ventricle: Fully Automated Segmentation Based on Spatiotemporal Continuity and Myocardium Information in Cine Cardiac Magnetic Resonance Imaging (LV-FAST)
【24h】

Left Ventricle: Fully Automated Segmentation Based on Spatiotemporal Continuity and Myocardium Information in Cine Cardiac Magnetic Resonance Imaging (LV-FAST)

机译:左心室:基于时空连续性和心肌信息的全自动分割在心脏磁共振成像(LV-FAST)中

获取原文

摘要

CMR quantification of LV chamber volumes typically and manually defines the basal-most LV, which adds processing time and user-dependence. This study developed an LV segmentation method that is fully automated based on the spatiotemporal continuity of the LV (LV-FAST). An iteratively decreasing threshold region growing approach was used first from the midventricle to the apex, until the LV area and shape discontinued, and then from midventricle to the base, until less than 50% of the myocardium circumference was observable. Region growth was constrained by LV spatiotemporal continuity to improve robustness of apical and basal segmentations. The LV-FAST method was compared with manual tracing on cardiac cine MRI data of 45 consecutive patients. Of the 45 patients, LV-FAST and manual selection identified the same apical slices at both ED and ES and the same basal slices at both ED and ES in 38, 38, 38, and 41 cases, respectively, and their measurements agreed within-1.6±8.7 mL,-1.4±7.8 mL, and1.0±5.8%for EDV, ESV, and EF, respectively. LV-FAST allowed LV volume-time course quantitatively measured within 3 seconds on a standard desktop computer, which is fast and accurate for processing the cine volumetric cardiac MRI data, and enables LV filling course quantification over the cardiac cycle.
机译:左室容积的CMR量化通常并手动定义了最基础的左室,这增加了处理时间和用户依赖性。本研究开发了一种基于LV的时空连续性的全自动LV分割方法(LV-FAST)。首先从心室中部到顶点使用迭代递减的阈值区域增长方法,直到左心室面积和形状不连续,然后从心室中部到基部,直到观察到不到50%的心肌周长。 LV时空连续性限制了区域的增长,以提高根尖和基底节段的鲁棒性。将LV-FAST方法与手动跟踪对连续45例患者的心脏MRI数据进行比较。在45例患者中,分别在38、38、38和41例患者中,LV-FAST和手动选择在ED和ES处识别出相同的根尖切片,在ED和ES处识别出相同的根尖切片,并且其测量结果符合以下条件: EDV,ESV和EF分别为1.6±8.7 mL,-1.4±7.8 mL和1.0±5.8%。 LV-FAST允许在标准台式计算机上在3秒钟内定量测量LV体积-时间过程,这对于处理电影体积的心脏MRI数据是快速而准确的,并且可以在整个心动周期内量化LV填充过程。

著录项