Evaluating the Therapeutic Dose Distribution of Intensity-Modulated Radiation Therapy for Head and Neck with Cone-Beam Computed Tomography Image: A Methodological Study

机译:锥形束计算机断层扫描图像评估头颈部调强放射疗法的治疗剂量分布:方法学研究

获取原文

摘要

An approximate correction method for the CT value-electron density curve of CBCT was established, through comparison and fitting with FBCT images, and applied to evaluate the therapeutic dose of IMRT. The precision of using CBCT for plan calculation was validated by comparing the dose distribution between CBCT- and FBCT-based IMRT plans. Also setup deviations were simulated to evaluate the ability of the CBCT-based calculation for detecting the dose errors caused by positioning deviation. The gamma comparison between CBCT- and FBCT-based dose computations showed that the pass rates of (2%, 2 mm) criteria were better than 97.60 ± 0.83% and 97.74 ± 2.08% in the phantom and 10 NPC cases. When setup deviation was introduced into CBCT-based dose calculation, the gamma pass rate significantly decreased while the volumetric doses of the targets and some normal organs exhibited different changes compared to the original plan. Our results validated the above CT value-electron density correction which reduced the difference between CBCT- and FBCT-based IMRT plan calculation for NPC to less than 2%. Online CBCT-based dose calculation can be used to reflect and evaluate the dose distribution discrepancy caused by setup deviation and structure changes during the treatment, ensuring more effective quality control of IMRT treatment.
机译:通过与FBCT图像的比较和拟合,建立了CBCT的CT值-电子密度曲线的近似校正方法,并用于评估IMRT的治疗剂量。通过比较基于CBCT和基于FBCT的IMRT计划之间的剂量分布,验证了使用CBCT进行计划计算的准确性。还模拟了安装偏差,以评估基于CBCT的计算能力,以检测由定位偏差引起的剂量误差。基于CBCT和FBCT的剂量计算之间的伽马比较显示,在幻影和10个NPC病例中,(2%,2 mm)标准的通过率优于97.60±0.83%和97.74±2.08%。当将建立偏差引入基于CBCT的剂量计算中时,与原始计划相比,伽玛通过率显着降低,而目标和一些正常器官的体积剂量显示出不同的变化。我们的结果验证了上述CT值-电子密度校正,该校正将NPC的基于CBCT和基于FBCT的IMRT计划计算之间的差异减小到不足2%。基于CBCT的在线剂量计算可用于反映和评估由治疗过程中设置偏差和结构变化引起的剂量分布差异,从而确保对IMRT治疗进行更有效的质量控制。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号