Long term response stability of a well-type ionization chamber used in calibration of high dose rate brachytherapy sources

机译:井型电离室的长期响应稳定性,用于高剂量率近距离放射治疗源的校准

获取原文
       

摘要

Well-type ionization chamber is often used to measure strength of brachytherapy sources. This study aims to check long term response stability of High Dose Rate (HDR)-1000 Plus well-type ionization chamber in terms of reference air kerma rate (RAKR) of a reference 137Cs brachytherapy source and recommend an optimum frequency of recalibration. An HDR-1000 Plus well-type ionization chamber, a reference 137Cs brachytherapy source (CDCSJ5), and a MAX-4000 electrometer were used in this study. The HDR-1000 Plus well-type chamber was calibrated in terms of reference air kerma rate by the Standards Laboratory of the International Atomic Energy Agency (IAEA), Vienna. The response of the chamber was verified at regular intervals over a period of eight years using the reference 137Cs source. All required correction factors were applied in the calculation of the RAKR of the 137Cs source. This study reveals that the response of the HDR-1000 Plus well-type chamber was well within ±0.5% for about three years after calibration/recalibration. However, it shows deviations larger than ±0.5% after three years of calibration/recalibration and the maximum variation in response of the chamber during an eight year period was 1.71%. The optimum frequency of recalibration of a high dose rate well-type chamber should be three years.Keywords: Brachytherapy, response stability, source calibration, well chamber
机译:井型电离室通常用于测量近距离放射治疗源的强度。这项研究旨在根据参考137Cs近距离放射治疗源的参考空气比释动能比(RAKR)来检查高剂量率(HDR)-1000 Plus良好型电离室的长期响应稳定性,并建议最佳的重新校准频率。本研究使用HDR-1000 Plus井型电离室,参考137Cs近距离放射治疗源(CDCSJ5)和MAX-4000静电计。 HDR-1000 Plus井型室是由维也纳国际原子能机构(IAEA)的标准实验室根据参考空气比释动能比进行校准的。使用参考137Cs放射源在八年内定期验证了反应室的响应。在计算137Cs源的RAKR时应用了所有必需的校正因子。这项研究表明,在校准/重新校准后的大约三年内,HDR-1000 Plus井型腔室的响应在±0.5%的范围内。但是,在三年的校准/重新校准后,其显示的偏差大于±0.5%,并且在八年期间内腔室响应的最大变化为1.71%。高剂量率井型腔室的最佳重新校准频率应为三年。关键词:近距离放射治疗,反应稳定性,源校准,腔室

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号