On-line digestion in a focused microwave-assisted oven for elements determination in orange juice by inductively coupled plasma optical emission spectrometry

机译:聚焦微波辅助烤箱中的在线消解,通过电感耦合等离子体发射光谱法测定橙汁中的元素

获取原文
   

获取外文期刊封面封底 >>

       

摘要

A flow injection system coupled to focused microwave-assisted oven was used for on-line orange juice sample digestion for determination of Ca, Cu, Fe, K, Mg, Mn, Na, P, and Zn by inductively coupled plasma optical emission spectrometry. The reactor coil was a PTFE tube (4.0 m long and 1.6 mm i.d.) positioned into the commercial glass tube of the focused microwave oven. Aliquots of 500 muL of sample and 1000 muL of reagent (80% v/v HNO3) were mixed in a confluence and carried out to the reactor coil by air carrier. The relative standard deviation for five replicates of sample was lower than 5.0%. Good recoveries varying from 91 to 111% were obtained for added concentrations of the interest elements. The results obtained using the proposed digestion system are in agreement with those obtained for total digestion at the 95% confidence level. With this on-line digestion system was possible to carry out 12 samples h-1, minimizing contamination, saving consumption of samples and reagent and low residue generation.
机译:流动注射系统与聚焦微波辅助烘箱相连,用于在线橙汁样品消解,通过电感耦合等离子体发射光谱法测定Ca,Cu,Fe,K,Mg,Mn,Na,P和Zn。反应器盘管是放置在聚焦微波炉的商用玻璃管中的PTFE管(长4.0 m,内径1.6 mm)。将等份的500 µL样品和1000 µL试剂(80%v / v HNO3)混合在一起,并通过空气载体送入反应器盘管。五个重复样品的相对标准偏差低于5.0%。对于所关注元素的添加浓度,获得了91%至111%的良好回收率。使用建议的消化系统获得的结果与以95%置信度进行总消化所获得的结果一致。通过这种在线消解系统,可以处理12个样品h-1,最大限度地减少污染,节省样品和试剂的消耗,并减少残留物的产生。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号