首页> 外文期刊>Journal of Science: Advanced Materials and Devices >Dielectric analysis of aqueous poly(l-glutamic acid) and poly-l-(glutamic acid4, Tyrosine1) solutions at high frequencies from capacitance measurements
【24h】

Dielectric analysis of aqueous poly(l-glutamic acid) and poly-l-(glutamic acid4, Tyrosine1) solutions at high frequencies from capacitance measurements

机译:水性聚( l l -(谷氨酸< ce:inf loc =“ post”> 4 ,酪氨酸 1 )在高频下通过电容测量解决方案

获取原文

摘要

A new parallel-plate capacitor fixture has been designed and successfully used to measure dielectric loss of polyelectrolyte solutions with volumes as low as droplets of 13–26?μL. It is particularly useful when studying polypeptides that are either high-cost or can be synthesized only in limited quantities. The ease with which the fixture can be used to obtain preliminary dielectric loss data yields savings in time and cost. In this study capacitance measurements were performed in a wide range of frequencies between 1 and 800?MHz using an Agilent 4191RF Impedance Analyzer. Accuracy of measurements was carefully examined through a comparison of measured conductivity of 1M NaCl against Stogryn's equation for conductivity. A 0.3% difference between the experimentally measured and theoretically calculated results has been found, demonstrating the validity of the proposed analysis method.
机译:设计了一种新的平行板电容器夹具,并成功地用于测量体积小至13–26?L的液滴的聚电解质溶液的介电损耗。当研究成本高或只能有限数量合成的多肽时,它特别有用。可以使用夹具轻松获得初步的介电损耗数据,从而节省了时间和成本。在这项研究中,使用Agilent 4191RF阻抗分析仪在1至800?MHz的较宽频率范围内进行了电容测量。通过将1M NaCl的电导率与Stogryn的电导率方程进行比较,仔细检查了测量的准确性。实验测量结果与理论计算结果之间存在0.3%的差异,证明了所提出分析方法的有效性。