Validity and reliability of a system to measure passive tissue characteristics of the lumbar region during trunk lateral bending in people with and people without low back pain

机译:在腰背痛患者和没有腰背痛患者中测量躯干横向弯曲过程中腰部区域被动组织特征的系统的有效性和可靠性

获取原文
           

摘要

The current study examined the validity and reliability of a new system that was developed to measure lumbar region passive stiffness and end range of motion during a trunk lateral bending movement in vivo. Variables measured included force, end range lumbar region motion, torque, lumbar region stiffness, and passive elastic energy. Validity of the force measurements was examined using standard weights. Validity of lumbar region angle measurements was examined using an instrumented trunk with an electrogoniometer. Reliability of the measurements between trials within a session was examined in a sample of 50 people (25 men, 25 women; mean +/ standard deviation age = 30.7 +/ 8.9 yr); 31 people reported a history of chronic or recurrent low back pain (LBP) and 19 reported no prior history of LBP. The end range lumbar region motion and force measurements demonstrated an excellent linear relationship with the criterion standard measures. Average error between the criterion standard and observed measurements was minimal for all measurements. For reliability testing, the majority of intraclass correlation coefficient values were >0.75. The validity and reliability of the current system are sufficient to examine lumbar region stiffness and end range of motion in people with and people without LBP.
机译:当前的研究检查了一种新系统的有效性和可靠性,该新系统用于测量体内躯干横向弯曲运动过程中腰椎区域的被动僵硬度和运动的最终范围。测得的变量包括力,腰部区域运动范围,扭矩,腰部区域刚度和被动弹性能。使用标准砝码检查力测量的有效性。腰部角度测量的有效性使用带电测角仪的躯干检查。在一个疗程中,两次试验之间的测量结果的可靠性在50人的样本中进行了检查(25位男性,25位女性;平均+ /标准偏差年龄= 30.7 + / 8.9岁); 31人报告有慢性或复发性下腰痛(LBP)的病史,19人报告无LBP的既往病史。末端范围的腰椎区域运动和力量测量结果与标准标准测量结果显示出极好的线性关系。对于所有测量,标准标准与观察到的测量之间的平均误差最小。对于可靠性测试,大多数组内相关系数值均大于0.75。当前系统的有效性和可靠性足以检查有LBP和无LBP的人的腰椎区域刚度和运动的最终范围。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号