Liquid waveguide spectrophotometric measurements of arsenate and particulate arsenic, as well as phosphate and particulate phosphorus, in seawater

机译:液体波导分光光度法测定海水中砷和微粒砷以及磷酸盐和微粒磷

获取原文
           

摘要

Sensitive methods for the determination of arsenate and particulate arsenic (PAs), as well as phosphate and particulate phosphorus (PP), in seawater are described. The method for arsenate and phosphate was established by applying automated liquid waveguide spectrophotometry. Because the reaction time for the formation of the arsenate-molybdate complex is longer than that for the phosphate-molybdate complex, a long Teflon tube submerged in a heating bath was installed in the conventional phosphate flow system. The arsenate was quantified as the difference between absorbances of molybdenum blue dyes with (only phosphate) and without (phosphate?+?arsenate) arsenate reduction treatment. Contamination was observed in the reagent for arsenate reduction and must be corrected. Linear dynamic ranges up to 1000?nM were confirmed for arsenate and phosphate. The detection limits for arsenate and phosphate were 5 and 4?nM, respectively. Freezing was a reliable sample preservation technique for both arsenate and phosphate. The method for PAs and PP was established by combining conventional persulfate oxidation of PP and the automated liquid waveguide spectrophotometry of arsenate and phosphate. The digestion efficiencies of organic As analogs were >93%. Contamination in the glass fiber filter was negligible. Field tests confirmed that the coefficients of variation (CVs) of 10–19?nM arsenate and 4–151?nM phosphate were 7–20 and 1–25%, respectively, while the CVs of 0.9?nM PAs and 10.2?nM PP were 11 and 4%, respectively.
机译:描述了测定海水中砷和微粒砷(PAs)以及磷酸盐和微粒磷(PP)的灵敏方法。采用自动液体波导分光光度法建立了砷酸盐和磷酸盐的测定方法。因为形成砷酸盐-钼酸盐络合物的反应时间比磷酸盐-钼酸盐络合物的反应时间长,所以在传统的磷酸盐流动系统中安装了浸没在加热浴中的长特氟隆管。砷酸盐的定量是用(仅磷酸盐)和不经(磷酸盐+砷酸)砷酸盐还原处理的钼蓝染料的吸光度之差。在试剂中观察到了污染,砷被还原,必须予以纠正。砷酸根和磷酸盐的线性动态范围高达1000?nM。砷酸盐和磷酸盐的检出限分别为5和4?nM。冷冻是砷酸盐和磷酸盐的可靠样品保存技术。通过将传统的PP过硫酸盐氧化与砷酸盐和磷酸盐的自动液体波导分光光度法相结合,建立了PA和PP的方法。有机砷类似物的消化效率> 93%。玻璃纤维过滤器中的污染可以忽略不计。现场测试证实,10–19?nM砷酸盐和4–151?nM磷酸盐的变异系数(CV)分别为7–​​20和1–25%,而0.9?nM PA和10.2?nM PP的CVs分别是11%和4%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号