A magnetic resonance imaging (MRI) guided stereotactic system was developed to provide veterinarians a method to accomplish minimally invasive stereotactic brain biopsies and procedures involving the cerebrum in canines. While MR-guided procedures are prevalent for humans, they are less common in animal practices. The system was designed to minimize fabrication costs in an effort to make such procedures more accessible in the veterinary field. A frame constrained the head without the need for punctures and supported registration and guidance attachments. Location data for registration and relevant structures were selected by the clinician, and a reverse kinematic analysis program generated the settings of the stereotactic arch to guide a needle to the desired location. Phantom experiments and three cadaver trials showed an average targeting error of <3 mm using the system.
机译:开发了磁共振成像(MRI)引导的立体定向系统,以为兽医提供一种完成微创立体定向脑活检和犬类涉及大脑过程的方法。尽管MR指导的程序在人类中很普遍,但在动物实践中却很少见。该系统旨在最大程度地降低制造成本,以使此类程序在兽医领域更容易获得。框架可以约束头部,而无需穿刺以及支撑的套准和引导附件。临床医生选择用于注册和相关结构的位置数据,然后进行反向运动分析程序生成立体定向弓的设置,以将针引导到所需位置。幻影实验和三个尸体试验显示,使用该系统的平均瞄准误差小于3mm。