首页> 外文期刊>Journal of Korean Neurosurgical Society >The BioFlex System as a Dynamic Stabilization Device : Does It Preserve Lumbar Motion?
【24h】

The BioFlex System as a Dynamic Stabilization Device : Does It Preserve Lumbar Motion?

机译:BioFlex系统作为动态稳定装置:它能保持腰部运动吗?

获取原文
           

摘要

Objective This study examines whether functional motion is present at one or more years after Bioflex System placement. BioFlex System is a flexible rod system which has been used to preserve motion at the area of implantation. There has not been a scientific study showing how much motion is preserved after implantation. Methods A total of 12 consecutive patients underwent posterior dynamic stabilization using the BioFlex System. Six patients were treated using a L3-4-5 construct and other six patients using a L4-5-S1 construct. Follow-up ranged from 12 to 33 months and standing neutral lateral, extension, flexion and posteroanterior (PA) radiographs were obtained at 3, 6, 9, and 12 months and at more than 12 months postoperatively. Range of motion (ROM), whole lumbar lordosis, and ROMs of motion segments from L2 to S1 were determined. Results Patients with a L3-4-5 construct demonstrated a decrease in mean ROM for whole lumbar decreased from 40.08 to 30.77. Mean ROM for L3-4 (6.12 to 2.20) and L4-5 (6.55 to 1.67) also decreased after one year. Patients with a L4-5-S1 construct demonstrated L4-5 (8.75 to 2.70) and L5-S1 (9.97 to 3.25) decrease of mean ROM at one year postoperatively. Lumbar lordosis was preserved at both L3-4-5 and L4-5-S1 constructs. Clinical results showed significant improvements in both study groups. Conclusion The present study provides preliminary information regarding the BioFlex motion preservation system. We conclude that the BioFlex System preserves functional motion to some degree at instrumented levels. However, although total lumbar lordosis was preserved, ROMs at implantation segments were lower than preoperative values.
机译:目的这项研究检查了Bioflex系统放置后一年或一年以上是否存在功能性运动。 BioFlex System是一种柔性杆系统,已用于在植入区域保持运动。尚未有科学研究表明植入后保留了多少运动。方法使用BioFlex系统对12例连续患者进行后动态稳定治疗。使用L3-4-5构建体治疗了6名患者,使用L4-5-S1构建体治疗了其他6名患者。随访时间为12到33个月,并在术后3、6、9和12个月以及超过12个月时获得了站立的中立侧位,伸展,屈曲和后前位(PA)的X线照片。确定了运动范围(ROM),整个腰椎前凸和从L2到S1的运动节段的ROM。结果L3-4-5结构患者的平均ROM降低,整个腰椎从40.08降低到30.77。一年后,L3-4(6.12至2.20)和L4-5(6.55至1.67)的平均ROM也降低了。具有L4-5-S1构建体的患者在术后一年时平均ROM降低了L4-5(8.75至2.70)和L5-S1(9.97至3.25)。腰椎前凸在L3-4-5和L4-5-S1构造中均得到保留。临床结果显示,两个研究组均有显着改善。结论本研究提供了有关BioFlex运动保存系统的初步信息。我们得出的结论是,BioFlex系统在仪器水平上可以在某种程度上保留功能性运动。然而,尽管保留了全部腰椎前凸,但植入段的ROM低于术前值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号